Performance Verification
3-10 ME7838G 0.8 mm Coaxial System Verification Procedure
VectorStar ME7838 Series MM
PN: 10410-00306 Rev. F
3-21
10.
If the verification fails, check the connectors of the test ports on the MA25400A Millimeter-Wave
modules, calibration components, and the verification standards for damage, cleanliness, and proper
connection and torquing. Also check connections of the interconnect RF/IF coaxial cables and their phase
stability. These are the most common causes for verification failures.
Note
Since different calibration methods are required to cover the 145 GHz range (SOLT for low band and
SSST for high band), the verification is done in two steps — Low band and High band with the
breakpoint set to 80 GHz.