2-7 Low-Band Calibration
Using the PVS Software with the ME7838A/AX/E/EX
2-20
PN: 10410-00784 Rev. A
3656C W1 Cal/Ver Kit & 2300-616-R PVS UG
3.
The VNA measures the load device data, then the
VNA Calibration: Step 2 of 5
dialog box appears.
a.
Connect the Male Cal Kit Broadband Load Device to VNA Port 1.
b.
Connect the Female Broadband Load Device to VNA Port 2.
c.
When ready to proceed, click
OK
.
4.
The VNA measures the broadband load device data, then the
VNA Calibration: Step 3 of 5
dialog box
appears.
a.
Connect the Male Cal Kit Open to VNA Port 1.
b.
Connect the Female Short #1 (2.02 mm) to VNA Port 2.
c.
When ready to proceed, click
OK
.
5.
The VNA measures the open/short device data, then the
VNA Calibration: Step 4 of 5
dialog box appears.
a.
Connect the Male Short #1 (2.02 mm) to VNA Port 1.
b.
Connect the Female Open to VNA Port 2.
c.
When ready to proceed, click
OK
.
Figure 2-26.
VNA Calibration: Step 2 of 5 Dialog
Figure 2-27.
VNA Calibration: Step 3 of 5 Dialog
Figure 2-28.
VNA Calibration: Step 4 of 5 Dialog