Using the PVS Software with the ME7828A
4-13 Restart for Alternate Frequency Band
3656C W1 Cal/Ver Kit & 2300-616-R PVS UG
PN: 10410-00784 Rev. A
4-35
4-13 Restart for Alternate Frequency Band
After the
Matched Thru
and
Mismatched Thru
tests have been successfully completed and the data saved, it is
necessary to restart the program and test the other frequency band. For example, if you tested for
40 MHz to
67 GHz Frequency Range
on your first calibration pass, on the second pass, select the
67 GHz to 110 GHz
Frequency Range
.
Procedure
1.
Ensure that the verification results of the previous test have been saved to the
C:\mmdscvc\Data
.
2.
To perform the verification tests for the second frequency range, on the
System Verification Software Main
Menu - Main Menu Tab,
select the
Restart
button. The main menu is described in
Interface – Main Menu Tab” on page 4-23
3.
The
Choose The Restart Method
dialog box appears as shown below in
.
4.
If you selected
Full Restart
, the calibration process starts again and the procedure steps are the same as
the procedures described previously starting in
Section 4-5, "Running the Verification Software"
5.
The application will go through the process again.
6.
Complete the calibration and verification of the devices for the second frequency range.
7.
Save the second set of calibration data and verification results to
C:\mmdcsvc\Data
.
8.
This series of automated tests has verified that the VectorStar ME7828A Broadband S-parameter
measurement accuracy meets factory specifications and ends the calibration and verification procedure.
Figure 4-49.
Choose The Restart Method Dialog
Note
When using the Restart Menu, the system information from the fist pass is retained when cycling
through the second pass.