4-8 High-Band Calibration
Using the PVS Software with the ME7828A
4-22
PN: 10410-00784 Rev. A
3656C W1 Cal/Ver Kit & 2300-616-R PVS UG
6.
The
VNA Calibration: Step 5 of 5
dialog box appears.
a.
Remove the offset Shorts from Port 1 and Port 2.
b.
Slide the two T-R modules towards each other and connect the VNA Port 1 (Female) to VNA Port 2
(Male) directly.
c.
When ready to proceed, click
OK
.
7.
The program execution continues to
Section 4-9 “Application Interface – Main Menu Tab” on page 4-23
.
Figure 4-34.
VNA Calibration: Step 5 of 5 Dialog
Summary of Contents for 3656C W1