Using the PVS Software with the ME7828A
4-5 Running the Verification Software
3656C W1 Cal/Ver Kit & 2300-616-R PVS UG
PN: 10410-00784 Rev. A
4-5
4-5
Running the Verification Software
After the VNA and PC have been configured as described above, you are ready to run the program and perform
the calibration/verification. Make sure you have the following information available:
•
Serial number of the 3656C Calibration/Verification Kit
•
Serial number of the 3739A Broadband Test Set
•
Model number of the LO Source (upper signal generator, usually an MG37022A)
•
Serial number of the LO Source (from the upper signal generator’s rear panel)
•
Model number of the RF Source (lower signal generator, usually an MG37022A)
•
Serial number of the RF Source (from the lower signal generator’s rear panel)
The application process flow for a full verification of an ME7828A is as follows:
1.
Choose your system configuration for Calibration/Verification kit type and system model number.
2.
Measure Low Band
a.
Calibrate the low frequency band
b.
Measure verification devices for low frequency band
c.
Save verification data
d.
Restart application for high frequency band
3.
Measure High Band
a.
Calibrate the high frequency band
b.
Measure verification devices for the high frequency band
c.
Save verification data
d.
Exit program.
4Note: the process can do the low band and high band in any order.
1.
With the equipment and software configured as described in
“Configuring the Hardware” on page 4-1
,
turn on the computer and allow it to boot up to Windows.
2.
Locate the
W1 Verification Software
icon (shown below) on the desktop and double-click on it.
Note
Low-band (step 2) and high-band (step 3) calibration and verification can be performed in any order.