6.2.28 13113
Device sensor signal level is too high. The possible reason of too high signal level is incorrectly put
device sensor.
Check the sensor and repeat the test.
6.2.29 13114
The signal level of two sensors is too low. The possible reason of too low signal level is incorrectly put
device sensors or opaque nail polish.
Eliminate the possible causes and repeat the test.
6.2.30 13115
The total duration of the Occlusion test has exceeded the time set, and the test was automatically
stopped. The test program is uncompleted and the test results are unreliable.
The possible reason is the late start of Occlusion test, too long Occlusion phase, uncomplete Occlusion
or lack of it.
6.2.31 13116
The Occlusion Test is finished, but accumulated data is not enough for further analysis as Occlusion
phase was not long enough. The test results should be treated as estimates.
To save test results in the database click "Save" toolbar button of Measurements window.
6.2.32 13119
Device memory card failed. Try to switch off and restart the device.
6.2.33 13121
The message says that the user specified has been already registered in computer's database. Test
results saved in device will be transferred to the already registered user.
6.2.34 13123
The message tells about the number of test results downloaded from device to computer. Only those
tests will be downloaded that have not been downloaded before.
6.2.35 "New users are found in the memory" Dialogue
When connecting the stand-alone device AngioScan-01P to computer, AngioScan program analyses the
data stored in memory of stand-alone device. If there are records of users not registered in computer's
database, AngioScan offers to register them. During the registration, test data is copied from device to
computer's memory, that further allows to operate the information more comfortably with the help of
patients database dialogue.
To refuse from the immediate registration click "Not now" button. Further, it will be possible to register
users and copy test results via "Settings" -> "AngioScan Settings" -> "Stand-alone device" tab, as well
as via patients database dialogue.
6.2.36 "Choose registration method" Dialogue
Here one can choose which method of registration in computer's database of users registered in stand-
alone device. The usual registration is preferable, as it allows to enter user's name and surname, that will
be displayed in database window.
Summary of Contents for AngioScan-01
Page 1: ......
Page 5: ...Chapter I...
Page 8: ...4 2014 AngioScan Electronics LLC Introduction...
Page 9: ...Introduction 5 2014 AngioScan Electronics LLC 1 7 Certificate in Regulatory Compliance...
Page 10: ...6 2014 AngioScan Electronics LLC Introduction 1 8 FSR Registration Certificate...
Page 11: ...Introduction 7 2014 AngioScan Electronics LLC 1 9 Patent Certificate for Invention...
Page 12: ...Chapter II...
Page 15: ...Chapter III...
Page 29: ...Quick Start 25 2014 AngioScan Electronics LLC...
Page 30: ...Chapter IV...
Page 58: ...Chapter V...
Page 109: ...Chapter VI...