background image

8 Two Wire

Four Wire

.

two wire test lead

connections always use the far left input

jacks

V

V

mode switch

in the 2 Wire

position

should be used for

all voltage, current and simple ohms

include all in-series wiring resistance

4 Wire Kelvin

.

input or

Kelvin mea-

surement test lead connection inputs and mode

selection switch For

both

‘red’ and

‘white’, which

are shown as a dotted line below in item # 8.
The

must also be

. Two wire mode

mea-

surements. 2 wire electrical resistance measure-

ments
from the meter out to the "resistance under test"

and back to the meter (connection/contact resis-

tance, test lead length, harness wiring length and
more etc.). The "Four Wire" switch position is

only for 4 Wire Kelvin resistance measurements

only. Basically the 4 Wire Kelvin connections
minimize the impact of test lead length resis-

tance and test lead contact resistance offsets. A
discussion on

may be found

further in this manual.

The

High

Low

9. Zero

(for the lower 3 re-

sistance ranges)

A.)

B.)

.0000 without a minus sign.

- The AMPTEC 601ES Explosive

Safety DMM has a precision ten turn zero pot.

zero knob is turned to "null-out" most

2 wire test lead resistance

. The zero adjust procedure in-

volves:

Short the test leads at the “UUT” end to-

gether while connected to the AMPTEC 601ES
DMM then

adjusting the zero knob until the

display shows

If

you get a minus sign while adjusting the zero,
you have over adjusted the zero knob. If this is
the case, turn the knob back the opposite direc-
tion until the minus sign is no longer displayed.
Once the zero is adjusted you can connect your
test leads to the "device under test". Re-zeroing
(2 wire mode) will be necessary only if you
change test leads. Do not force the zero knob
when turning it, especially once the stops have
been reached. Forcing the zero knob past the
stops will severely damage the precision of the
instrument and violate the instruments warranty.

( 2, 20 ,and 200 ohm ranges)

10. Case/Chassis Ground Jack

(not tied to earth ground)

A.)

B.)

provides the

user access to the AMPTEC 601ES DMM’s
Faraday Cage and Chassis ground. In some ap-
plications (i.e. anti-static) the AMPTEC 601ES
DMM user may want to connect your "In-House
local Earth Ground" to the AMPTEC 601ES
DMM chassis ground. All appropriate internal
grounds of the AMPTEC 601ES DMM are con-
nected to the front panel chassis ground jack.
The unit's case ground jack accepts banana jacks,
spade lugs and even bare wires for making an
earth to case ground connection. In other appli-
cations, the AMPTEC 601ES DMM user may
want the meter to be floating from the resistance
under test

and the

case ground jack may be left disconnected or
ignored.

The AMPTEC 601ES DMM has four differ-

ent ranges for measuring AC/DC voltage. The
meter starts at the 600VRMS test level. Should
the user determine there is virtually no measur-
able voltage on the high (600V) range then
select the next lower range (200V).

Rotating the knob to the next (clockwise)

range selects the 200V True RMS AC-DC
Voltage measurement function (200 Volts
fullscale max.). This DC coupled voltage mea-
surement function will therefore display any DC
Voltage - up to 200 V max. This voltage mea-
surement function also measures True RMS AC
Voltage including distorted waveforms (band-
width to 100 Hz). The process of stepping to the
next progressively lower voltage range would
normally continue until the lowest range of 2.0
VDC- AC RMS. The 2.0 V range display

11. AC-DC Voltage VRMS Measurement

STRAY CURRENT

PRESS TO MEASURE

2.0 KOHM

20 OHM

200 OHM

2.0 OHM

20 KOHM

FUNCTION / RANGE

2 AMP

AC/DC

600VRMS

200V

AC/DC

20V

AC/DC

2.0V

AC/DC

+ HIGH

CASE

GROUND

LOW

V

V

LO

V

HI

I

2 Wire

4 Wire

4 wire Kelvin ohms

uses all 4 terminals

+ HIGH

LOW

V

V

LO

V

HI

I

Sect. D - Operation, Functional Self-Test and Use

12

Summary of Contents for 601ES

Page 1: ...RESEARCH AMPTEC CALIBRATED MODEL SER NO DATE DUE Tamper ProofSeal Sticker Calibrationvoidifremoved 601ES 601ES 503 05APR2005 05APR2006 I press hold 2 0 KOHM 20 OHM 200 OHM 2 0 OHM 20 KOHM FUNCTION RA...

Page 2: ...cifications C 1 Available Accessories and Options D 1 General D 2 Front Panel Features and Operation D 3 Complete 601ES Explosive Safety DMM Front Panel Diagram D 4 4 Wire Resistance Measurement 601ES...

Page 3: ...Procedure F 3 1 Zero Offset Adjustment F 3 2 Full Scale Adjustment F 3 3 2 0 Ohm Range Adjustments F 3 4 Verification of Each Ranges Accuracy F 3 5 AC DC Voltage Adjustments F 3 6 AC DC Current Adjust...

Page 4: ...osive Safety Igniter Circuit Test industry must be held down for 2 seconds The AMPTEC 601ES DMM Safety Meter has a circuit that indicates when it is time To replace the batteries They can be replaced...

Page 5: ...ince Nickel Metal Hydrides are re chargeable they offer a battery life cycle a hundred times greater than any alkaline Be sure to charge the Nickel Metal Hydride batteries when taken out of their orig...

Page 6: ...around the battery holder and secure Cut excess plastic Next wrap the black and red wire around the battery holder and place the holder into the compartment Lastly replace lid and screw in screws to...

Page 7: ...ange Accuracy 0 50 of range DC or AC RMS range Accuracy 0 50 of range DC or AC RMS range Accuracy 0 50 of range 2 0 Amperes DC or AC RMS range Accuracy 0 50 of range 2 0 Ohm Range thru 20 KOhm ranges...

Page 8: ...e point of the probe tip connection eliminating 98 of the cable resistance Gold plated tips and banana jack ends provide an optimal low thermal emf connection gold plated single points gold plated gol...

Page 9: ...banana plugs at one end The other end is terminated with an alligator clip red and black These alligator clips will measure 2 wire method properly with either upper or lower jaw connected to the condu...

Page 10: ...once the access screws have been removed for calibration These handles are for carrying the 601ES DMM around when the lid is removed They are strictly for calibration access he main Power switch to tu...

Page 11: ...M is factory set using internal PCB jumpers to automatically turn off after 30 minutes of inactivity no range or function change to conserve battery power 5 Automatic Turn Off Protection Fuse Stray Cu...

Page 12: ...g the zero knob past the stops will severely damage the precision of the instrument and violate the instruments warranty 2 20 and 200 ohm ranges 10 Case Chassis Ground Jack not tied to earth ground A...

Page 13: ...s go in continuous decades from 2 0 Ohms to 20 Kilo Ohms The selection of the 2 0 Ohm lowest resistance range is accomplished by rotating the range knob Align indicator mark on knob with the desired r...

Page 14: ...ck to the AMPTEC 601ES DMM through the jack Depending upon the magnitude of the resistance offset some of the offset may be corrected zero ed out for in the AMPTEC 601ES DMM by adjusting the zero knob...

Page 15: ...g Circuitry fatigued adapter wire plug wire splice loose the fewer the connections the fewer the problems The LCD display will indicate when it is time to replace the batteries LO BAT LO BAT The The A...

Page 16: ...atsoever If you turn on display does not come on Explosive Safety DMM m AMPTEC Explosive Safety DMM and the If the AMPTEC 601ES Explosive Safety DMM exhibits problems that cannot be eliminated by revi...

Page 17: ...only when the integrator is de integrating At the end of a conversion the data remaining in the results counter is latched decoded and multiplexed to the liquid crystal display The analog section blo...

Page 18: ...n also turn the unit off the same way after a certain time period has elapsed This is to save battery life See FIG H 1 The micro can turn on TR201 TR209 using IC201 and I 202 These fets are used to se...

Page 19: ...y Out of Specification Section B No adjustments are needed Verify each function and range with the appropriate calibration standards to determine if the AMPTEC 601ES DMM meets its published accuracy s...

Page 20: ...PCB then later verify 3 4 5 6 7 8 A B C D F 3 6 AC DC Current Adjustments 1 the battery compartment When adjusting RV405 which is indicated as a negative sign Select the 20 Kohm range Connect the Kel...

Page 21: ...press to measure button and Adjust RV502 and repeat this step 5 dual button press operation until the AMPTEC 601ES reads 0050 A 5 milliamperes Return both the and the s to their standoff PCB mounts Re...

Reviews: