36
when performing routine calibrations for most gas
analyzers, especially trace oxygen.
Other plastic tubing such as: Teflon, Tygon, Nylon or
Polyethylene diffuse oxygen generally in the range of
2ppm to 100ppm of oxygen per 15' lengths. The
amount of diffusion depends on temperature, flow and
pressure. Previously, low level trace oxygen
measurement required stainless steel or other metallic
tubing which is much too cumbersome and prone to
failure for routine calibrations.
Available for all 1/4" swage type compression fittings.
6,500psig. Rating.
Calibration gas regulator
This gas nipple with o-ring seal eliminates common
leaks associated with imperfect metal to metal surfaces
that are common when mating high pressure gas
cylinders to pressure reducing regulators.
This inexpensive brass nipple and o-ring combination
provides a simple and reliable solution to the common
occurrence of leaks produced by metal to metal
connections.
This nipple will fit any pressure reducing regulator inlet
with a ¼" NPT inlet port.
CD and USB cable
AMI User Interface Program.
Allows complete access to all advanced features of the
210BR analyzer.
Contact AMI for information about any of these.