Configuration
|
5-13
Flow Meter Configuration
To configure the flow meter, select Flow Config from the Configuration Menu.
Figure 5-16. Flow Configuration screen.
Use the drop down menu to select the untis for use on the analyzer
- scfh
- lpm
- ccm
Use the keypad to enter the low flow alarm threshold in the appropriate units.
Flow readings are not pressure compensated, therefore, measurement
is subject to error if sample flow is not vented to the atmosphere. In
this case, a flow calibration should be completed See Flow Calibra
-
tion Section 6-11.