10-2 | Thermox WDG-V
RM
register
parameter
type
description
Id
1
PARID_SYSALARM
Int 32
0 - host is communicating with the sensor;
1 - host is not communicating with the sensor
Read Only
61
3
PARID_PASSCODE
Unsigned Int 32 User pascode 000000 through 999999
Writable
85
5
PARID_SUPERCODE
Unsigned Int 32 Supervisor passcode 000000 through 999999
Writable
86
7
PARID_OXYGEN
Float
Oxygen Concetration in Percentage
Read Only
1
9
PARID_CELLTEMP
Float
Cell Temperature in deg C
Read Only
5
11
PARID_CELLHDC
Float
Cell heater duty cycle
Read Only
15
13
PARID_CELLTEMPSP
Float
O2 Cell temperature setpoint in deg C
Writable
17
15
PARID_PROCALARM-
MASK
Unsigned Int 32 Process alarm bit mask
Read Only
37
17
PARID_COMMAND
Unsigned Int 16 Command to the analyzer
Writable
42
18
PARID_CMDRESP
Unsigned Int 16 Response to a command
Read Only
43
19
PARID_O2SPANGAS
Float
Oxygen span gas value in %
Writable
44
21
PARID_O2ZEROGAS
Float
Zero gas value in %
Writable
45
23
PARID_DATAVALID
Unsigned Int 16 Datavalid (1 if valid, 0 if not)
Read Only
48
24
PARID_STATE
Unsigned Int 16 Current state of the analyzer
Read Only 108
25
PARID_FLOWTESTVAL
Unsigned Int 32 Result of the last flow test
Read Only 129
27
PARID_RELAY3FUNC
Unsigned Int 16 Relay 3 Function (0-Disable, 1-O2)
Writable
150
28
PARID_RELAY3HI
Float
Relay 3 Hi threshold limit
Writable
151
30
PARID_RELAY3LO
Float
Relay 3 Lo threshold limit
Writable
152
32
PARID_RELAY3ENRGZ
Unsigned Int 16 Relay 3 energize/de-energize on alarm
(0-deenergize, 1- energize)
Writable
153
33
PARID_EVENTMASK1
Unsigned Int 32 Event Mask 1
Read Only 177
35
PARID_EVENTMASK2
Unsigned Int 32 Event Mask 2
Read Only 178
37
PARID_PRESSURE
Float
Process Pressure
Writable
185
39
PARID_FLOWRATE
Unsigned Int 16 Flow rate 0- scfh, 1 - lpm, 2 - ccm
Writable
227
40
PARID_FLOWCONV
Float
Flow displayed in the units selected via flow
rate (reg 39)
Read Only 229
42
PARID_SETFLOWALARM Float
Sets the low flow alarm setpoint using the
units selected by flow rate (reg 39)
Writable
230
Summary of Contents for Thermox WDG-VRM
Page 1: ...Thermox WDG VRM User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 9000 166 VE REV H...
Page 13: ...System Overview 1 3 Figure 1 1 WDG VRM Schematic...
Page 24: ...2 4 Thermox WDG VRM This page intentionally left blank...
Page 27: ...Installation 3 3 Figure 3 1 WDG VRM front view Figure 3 2 WDG VRM mounting dimensions...
Page 34: ...3 10 Thermox WDG VRM Figure 3 7a Customer I O Interconnecting WDG VRM...
Page 35: ...Installation 3 11 B DETAIL B SCALE 6 4 Figure 3 7c Customer I O Interconnecting WDG VRM...
Page 44: ...3 20 Thermox WDG VRM This page intentionally left blank...
Page 103: ...Customer Modbus Map 10 1 Customer Modbus Map...