Specifications
|
2-3
Calibration
:
Calibrate or verify calibration. Store last calibration and verification data. Se
-
lectable calibration gas run time and process recovery time. Timed automatic
calibration with optional remote calibration unit.
Environment
:
Ambient Temp: -25°C to 50°C
Relative Humidity: 10-90%, non-condensing
Max Altitude: 3000 meters
Enclosure
:
19” Rack mount, 4U
Indoor use only.
Power Requirements
:
115 VAC ±10%; 47-63 Hz., 740 VA max.
230 VAC ±10%; 47-63 Hz., 740 VA max.
Fuse Rating 4A, 250VAC, SLO-BLOW
Calibration Gas Requirements
:
Use calibration gases at 2-20 scfh; should match process
O
2
Zero Gas:
0.1-10% O
2
, balance N
2
O
2
Span Gas:
Minimum one decade above zero gas (10 times greater)