Calibration
and Troubleshooting
7-24
M370430-01 Rev F
Offset Calibration
Offset calibration of the power supply provides the best accuracy
in low-range input current.
To perform offset calibration:
1.
Set the voltage source attached to the non-isolated current
APG input to 10% of the nominal current level, in this case
0.400 V.
2.
Turn the output power on.
3.
Read the input voltage value from the voltmeter connected in
step 2 of the gain calibration.
4.
Type SCPI
input current calibration command with voltage
noted in step 3 as the parameter “data”.
For example, you would type
:CAL:INP:ANAL:CURR 0.43
.
After performing offset calibration, it is highly recommended that
you repeat gain calibration.
SCPI Command
[:]CALibration[<channel>]:INPut:ANALog:CURRent{<voltme
ter reading in volts>}
Summary of Contents for Sorensen XG 100-15
Page 2: ......
Page 93: ...Local Operation M370430 01 Rev F 3 47 3 Figure 3 5 Series Operation ...
Page 99: ...Local Operation M370430 01 Rev F 3 53 3 ...
Page 100: ...Local Operation 3 54 M370430 01 Rev F ...
Page 220: ...Communications Options 6 14 M370430 01 Rev F ...
Page 244: ...Communications Options 6 38 M370430 01 Rev F ...
Page 252: ...Communications Options 6 46 M370430 01 Rev F ...
Page 292: ......
Page 329: ...M370430 01 Rev F A 37 A APPENDIX A SCPI Command Reference ...
Page 330: ......
Page 368: ...APPENDIX D Rack Mount Options and Installation Instructions D 12 M370430 01 Rev F ...
Page 369: ...APPENDIX D Rack Mount Options and Installation Instructions M370430 01 Rev F D 13 ...
Page 370: ......
Page 395: ...Rack Mount Options and Installation Instructions M370430 01 Rev F E 25 ...
Page 396: ......
Page 399: ......