Local
Operation
3-14
M370430-01 Rev F
To set the voltage set point (VSET)
:
1.
Turn the 9-position mode control to the VOLTS position or
press the rotary Adjust/Enter control if the 9-position mode
control is already at the VOLTS position.
The voltage set point is blinking in the output voltage display. The
output current will be displayed in the output current display.
2.
Set the desired voltage value using the rotary Adjust/Enter
control.
Important:
If you can't get the set point to the desired level, you may need to
change the set points for OVP.
3.
Press the rotary Adjust/Enter control to transition to Fine
Adjust mode.
The set point will blink faster to indicate fine adjust tracking mode.
4.
Set the desired voltage value using the rotary Adjust/Enter
control.
5.
Once the desired value has been set, press the rotary
Adjust/Enter control to commit the setting.
Important:
The control circuits have been designed to allow you to set the
output voltage up to 105% over the model-rated maximum value. The power
supply will operate within these extended ranges, but full performance to
specification is not guaranteed.
Quick Tip Remote
operation uses these
SCPI commands. See
page 5-1.
The SCPI command (s) for these instructions are:
[[:]SOURce[<channel>]]:VOLTage[:LEVEl][:IMMediate]
[:AMPLitude]
Important:
The maximum and minimum setting values of the output voltage
are limited by the Over Voltage Protection and Under Voltage Protection
settings.
To set the current set point
(I
SET
)
:
1.
Turn the 9-position mode control to the AMPS position or
press the rotary Adjust/Enter control if the 9-position mode
control is already at the AMPS position.
The voltage set point is blinking in the output voltage display. The
output current will be displayed in the output current display.
2.
Set the desired current value using the rotary Adjust/Enter
control.
3.
Press the rotary Adjust/Enter control to transition to Fine
Adjust mode.
The set point blinks faster to indicate fine adjust tracking mode.
Summary of Contents for Sorensen XG 100-15
Page 2: ......
Page 93: ...Local Operation M370430 01 Rev F 3 47 3 Figure 3 5 Series Operation ...
Page 99: ...Local Operation M370430 01 Rev F 3 53 3 ...
Page 100: ...Local Operation 3 54 M370430 01 Rev F ...
Page 220: ...Communications Options 6 14 M370430 01 Rev F ...
Page 244: ...Communications Options 6 38 M370430 01 Rev F ...
Page 252: ...Communications Options 6 46 M370430 01 Rev F ...
Page 292: ......
Page 329: ...M370430 01 Rev F A 37 A APPENDIX A SCPI Command Reference ...
Page 330: ......
Page 368: ...APPENDIX D Rack Mount Options and Installation Instructions D 12 M370430 01 Rev F ...
Page 369: ...APPENDIX D Rack Mount Options and Installation Instructions M370430 01 Rev F D 13 ...
Page 370: ......
Page 395: ...Rack Mount Options and Installation Instructions M370430 01 Rev F E 25 ...
Page 396: ......
Page 399: ......