Calibration and Troubleshooting
5-10
TM-F2OP-C1XN-01
Troubleshooting for Operators
Use the checks in Table 5-1 to ensure that the
power supply is configured
and connected for normal operation.
Abbreviated References Used in Table
ACF
AC Fail
OTP
over temperature protection
OVP
over voltage protection
REM
remote mode
S/D shutdown
Table 5-1
User Diagnostics
Symptom
Check
Further Checks and Corrections
No output and the
display is blank.
Is input voltage within specified
range?
Connect to appropriate voltage
source.
See page 2–6.
Power switch ON?
Turn on power.
Internal circuit?
See your service technician.
No output but the
display turns on.
OVP LED turned on?
See page 3–7.
Front panel S/D LED turned on?
OTP LED turned on?
Current limit set to zero?
Voltage control set to zero?
REM LED turned on?
If using remote analog control, check
your analog programming source
(Chapter 4). If not, refer to your
digital interface manual.
Is front panel ACF LED turned on?
Connect unit to AC supply in
specified range. See page 2–6.
Internal circuit.
See your service technician.