Features and Specifications
1-8
TM-F2OP-C1XN-01
Table 1-3
Rear Panel J2 Connector Terminals and Functions
Connector
Reference
Name
Function
J2-1
VRMT
Remote Output Voltage
Programming Select
Selects remote output voltage
programming when jumpered to pin 3.
J2-2
IRMT
Remote Output Current Limit
Programming Select
Selects remote output current limit
programming when jumpered to pin 3.
J2-3
CNTL
GND
Control Ground
Control ground.
J2-4
N/C
No connection
None.
J2-5
VPGM–
Voltage Program Signal Return
Return for voltage program signal.
J2-6
VPGM
Output Voltage Programming
Input
Input for voltage programming signals
from an analog device.
J2-7
IPGM–
Current Program Signal Return
Return for current program signal.
J2-8
IPGM
Output Current Limit
Programming Input
Input for current limit programming
signals from an analog device.
J2-9
VMON–
Voltage Monitor Signal Return
Return for voltage monitor signal.
J2-10
VMON
Output Voltage Monitor
Output for output voltage monitor
signal.
J2-11
IMON–
Current Monitor Signal Return
Return for current monitor signal.
J2-12
IMON
Output Current Monitor
Output for output current monitor
signal.
J2-13
N/C
No connection
None.
J2-14
S/D–
Shutdown Signal Return (–)
Return for shutdown signal.
J2-15
S/D
S/D Input (+)
Input for shutdown signal.
Summary of Contents for Sorensen XFR 100-28
Page 3: ...This page intentionally left blank ii...
Page 5: ...This page intentionally left blank iv...
Page 7: ...This page intentionally left blank vi...
Page 13: ...xii...
Page 15: ...xiv...
Page 25: ...1 10...
Page 67: ...3 18...
Page 73: ...Remote Operation 4 6 TM F2OP C1XN 01 Figure 4 1 Connecting Programming Sources to J2 Connector...
Page 75: ...4 8...
Page 87: ...5 12...