DLM-E 3kW & 4kW Series Power Supplies
Advanced Operation
Figure 4–1. J3 Connector
Pin No.
Function
Pin No.
Function
1
Remote Output Enable
1 = Enable, 0 = Disable
14
Remote Shutdown Input (+). Positive or
negative true logic selection with S1
2
Remote Shutdown Return (–)
15
+5 VDC Aux. Output
3
Remote OVP Programming Input
16
1 mA current source for OVP
Programming
4
Remote Programming Indicator
1 = Remote, 0 = Local
17
OVP Status Indicator
1 = OVP Shutdown, 0 = Normal
5
Operating Mode Indicator
1 = Volt mode, 0 = Current mode
18
Overtemp Shutdown Indicator
1 = OTP Shutdown, 0 = Normal
6
Status Indicator Return (–)
19
DC Voltage Monitor Output
7
Current Monitor Output
20
Remote/Local Voltage Control Select
8
N/C – for all models , except
16V, 22V models – Fault signal,
Active High, 10mA current source
21
1 mA current source for Voltage
Programming
9
Voltage Programming Input
22
1 mA current source for Current
Programming
10
Current Programming Input
23
Remote/Local Current Control Select
11 N/C
24 N/C
12
Programming/Monitor Return (–)
25
N/C
13 N/C
Table 4–1. J3 Connector – Program, Control, and Monitor Description
(D–subminiature 25–Pin Female)
4.2.2
Rear Panel DIP Switch
Switch S1 is located on the main printed circuit board and is able to be changed through the
rear panel of the power supply. The J3 connector is located on the unit's rear panel. See
Section 4.2.3
Resetting Rear Panel DIP Switch Settings
M362000-01 Rev E
4-3
Summary of Contents for Sorensen DLM-E
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...viii M362000 01 Rev E This page intentionally left blank...
Page 8: ...x M362000 01 Rev E This page intentionally left blank...
Page 10: ...xii M362000 01 Rev E This page intentionally left blank...
Page 12: ...xiv M362000 01 Rev E This page intentionally left blank...
Page 18: ......