Basic Operation
DLM-E 3kW & 4kW Series Power Supplies
Figure 3–1 provides a graphical representation of the relationships between these variables.
Constant Voltage Mode Operation
The power supply will operate in constant voltage mode whenever the load current I
L
is less
than
the current limit setting I
SET
, or: I
L
< I
SET
(Note: I
L
= V
SET
/ R
L
)
In constant voltage mode, the power supply maintains the output voltage at the selected value
(V
SET
) while the load current I
L
varies with the load requirements.
Constant Current Mode Operation
The power supply will operate in constant current mode whenever the load resistance is low
enough that the load current I
L
is greater than
the current limit setting I
SET
, or:
I
L
> I
SET
In constant current mode, the power supply maintains the output current at the selected value
(I
SET
) while the load voltage varies with the load requirements.
Figure 3–1. Operating Modes
Automatic Mode Crossover
This feature allows the power supply to automatically switch operating modes in response to
changing load requirements. If, for instance, the power supply was operating in Constant
Voltage Mode (I
L
< I
SET
), and the load changed so that the load current (I
L
) became
greater
than
the current limit setting (I
SET
), the power supply would automatically switch into Constant
Current Mode. If the additional load was subsequently removed so that the load current was
again
less than
the current limit setting, the supply would automatically return to Constant
Voltage Mode.
3-2
M362000-01 Rev E
Summary of Contents for Sorensen DLM-E
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...viii M362000 01 Rev E This page intentionally left blank...
Page 8: ...x M362000 01 Rev E This page intentionally left blank...
Page 10: ...xii M362000 01 Rev E This page intentionally left blank...
Page 12: ...xiv M362000 01 Rev E This page intentionally left blank...
Page 18: ......