PROPERTY OF AMETEK, INC.
–
ORTEC
®
ORTEC
®
EASY-NIM
™
928 Suite
932512A / 0114
24
●
Conversion Gain
— This sets the maximum channel number in the spectrum. If set to 16384, the energy scale
will be divided into 16384 channels. The conversion gain is entered in powers of 2 (e.g., 8192, 4096, 2048).
The up/down arrow buttons step through the valid settings.
●
Upper- and Lower-Level Discriminators
— The
Lower Level Disc
riminator sets the level of the lowest
amplitude pulse that will be stored. This level establishes a lower-level cutoff by channel number for ADC
conversions. The
Upper Level Disc
riminator sets the level of the highest amplitude pulse that will be stored.
This level establishes an upper-level cutoff by channel number for storage.
3.4.2.
About Tab
This tab (Figure 29) displays hardware and firmware information about the currently selected 928 as well as the data
Acquisition Start Time
and
Sample
description. In addition, the
Access
field shows whether the MCB is currently
locked with a password (see the password discussion in the MAESTRO user manual);
Read/Write
indicates the
instrument is unlocked;
Read Only
means it is locked.
Figure 29. About Tab.