3.3.4.3.
O
UTPUT
S
ETUP
The output setup fields are illustrated in Figure 16. Optionally define a
Friendly name
and select an output
Source
from the list.
3.3.4.4.
C
OUNTER
I
NPUT
S
ETUP
The counter input setup fields are shown in Figure 17.
Signal threshold (V)
— Adjustable from -1.6 V to +3 V in steps of 1.5 mV.
Polarity
— Select positive or negative signal polarity.
Impedance
— Choose 50 Ω or 1 kΩ input impedance.
Alarm threshold
— When count rate exceeds this value, the
CRM gauge
readout on the main screen flashes red. In
addition, if the
Play sound when alarm triggered
checkbox is marked on the setup screen, an audio alarm sounds. If
the output is configured to
CRM Alarm
, the output goes logic high.
Figure 16. Output Setup.
3.3.4.5.
S
AVING AND
L
OADING
S
ETTINGS
(.INI) F
ILES
The lower-right
Save to file
button allows you to create a.
.INI
file containing all current settings. The
Load settings
button allows you to recall the
.INI
file of your choice, and loads the settings into the currently selected 928.
Figure 18 shows the file-save dialog. Figure 19 shows an excerpt of a
.INI
file in Notepad.