PROPERTY OF AMETEK, INC.
–
ORTEC
®
ORTEC
®
EASY-NIM
™
928 Suite
932512A / 0114
56
7
Input 6 BNC
8
Timer Interval 1
9
Timer Interval 2
10
Software Group 1 Enable
11
Software Group 2 Enable
4
Counter input
Value
0
Corresponding Counter BNC Input
1
Connected to Overflow of Counter 2
2
Connected to Overflow of Counter 3
3
Connected to Overflow of Counter 4
4
Overflow of Timer 1
5
Overflow of Timer 2
6
ROI (counts all channels for which Region-of-Interest bit is set)
5
Signal threshold
Range -1.6 V to 3.0 V
6
Impedance
0 = 1 kΩ, 1 = 50 kΩ
7
Counter input signal edge
0 = rising edge, 1 = falling edge
8
Auto reset
0 = off, 1 = On
9
FIFO counts
Number of words in FIFO (this value is read only. Can only be used in
SHOW_COUNTER_CONFIG)
10 Acquisition status mask
(Read only. Can only be used in SHOW_COUNTER_CONFIG). Bit-mapped
flags indicate whether counter or timer has started counting.
Bit 0
Counter 1
Bit 1
Counter 2
Bit 2
Counter 3
Bit 3
Counter 4
Bit 4
Timer 1
Bit 5
Timer 2
11 Alarm threshold
counts per second
SET_OUTPUT_CONFIG n1, value
Set the output configuration where n1 is 1 and the corresponding output value is defined as follows:
0
Software Control
1
GND (0 V)
2
Logic high (5 V)
3
Counter 1 Overflow (Pulse)
4
Counter 2 Overflow (Pulse)
5
Counter 3 Overflow (Pulse)
6
Counter 4 Overflow (Pulse)
7
Timer 1 Overflow (Pulse)
8
Timer 2 Overflow (Pulse)
9
Counter 1 Overflow Level (must be Reset)
10
Counter 2 Overflow Level (must be Reset)
11
Counter 3 Overflow Level (must be Reset)
12
Counter 4 Overflow Level (must be Reset)
13
Timer 1 Overflow Level (must be Reset)
14
Timer 2 Overflow Level (must be Reset)
15
Timer 1 Interval
16
Timer 2 Interval
17
CRM Alarm 1
18
CRM Alarm 2