932512A / 0114
ORTEC
®
EASY-NIM
™
928 Suite
35
The MCB Clear Command signal (allows timer to be synchronized to MCB acquisitions).
Group n Reset (n = 1 or 2) resets all timers or counters and timers in the specified group.
●
Enable Input
Timer must be enabled to count. Enable can be configured to be controlled by the
MCB Active
signal, the
Software Control
signal, or the
Software Group n Enable
(for timer n, n = 1 or 2) signal, which
starts all timers for the specified group.
●
Overflow
Output provides a rising edge at the moment a timer reaches the value supplied for Preset Input.
The Timer will return to 0 at that moment and continue to count.
●
Time preset
Acquisition time specified by the number of repetitions of the currently selected
Time base
.
●
FIFO Readout Control
Timer FIFOs function the same as the counter FIFOs (see above).
4.3.
ELECTRICAL AND MECHANICAL
High-Speed USB
Standard connection via a supplied 3 m (10 ft) cable.
Indicators
●
ADC
Indicates activity for ADC-1.
●
Counter
Illuminated if any of the four counters is actively counting.
Power Required
+6 V, 350 mA; -6 V, 225 mA; +12 V, 100 mA; –12 V, 100 mA.
Weight
●
Net 0.9 kg (2 lb).
●
Shipping 2.25 kg (5 lb).
Dimensions
NIM-standard, single-wide 3.43 cm x 22.13 cm (1.35 in. x 8.714 in.) front panel per DOE/ER0457T.
4.4.
MCB SOFTWARE CONTROLS IN CONNECTIONS APPLICATIONS
●
ADC LLD
Computer controlled from 0 to 100% full scale.
●
ADC ULD
Computer controlled from 0 to 100% full scale.
●
Live Time Correction
Gedcke-Hale, ZDT.
●
Gate
Coincidence, Anticoincidence, Off.
●
Operating Systems
32-bit and 64-bit Microsoft Windows 7, Windows XP SP3.