background image

PROPERTY OF AMETEK, INC.

ORTEC

®

ORTEC

®

EASY-NIM

928 Suite

932512A / 0114

50

SHOW_MDA_COEF

Reports the current settings for coefficients a, b, and c used for the MDA calculation. See SET_MDA_COEF.
Example Responses:

MDA_COEF 000000000002.71 000000000000000 00000021.700001

Coefficient a = 2.71, b = 0, and c = 21.7.

SHOW_MDA_PRESET [PeakNumber]

Reports MDA preset PeakNumber. If PeakNumber is omitted, 0 is assumed.
Example Response:

MDA_PRES 0000000000025.7

SHOW_MODE

Reports the current mode of operation (PHA, List, or Sample [InSight Virtual Oscilloscope]). See also

SET_MODE_PHA and SET_MODE_SAMPLE.
Responses:

$FPHA<CR>

PHA mode.

$FLIS<CR>

List mode.

$FSAM<CR>

Sample mode (InSight).

SHOW_MODE_ZDT

Reports the current ZDT mode. See SET_MODE_ZDT.
Responses:

MODE_ZDT 00000

Mode 0 selected (normal and corrected).

MODE_ZDT 00001

Mode 1 selected (error and corrected).

SHOW_MONI_LABEL num

Returns with an ASCII string label designation for the state-of-health parameter for num.

SHOW_MONI_MAX

Returns with the number of state-of-health values available for viewing.

SHOW_MONI_VALUE num

Returns with an ASCII representation of the value for the state-of-health parameter.

SHOW_NEXT

Used in conjunction with the SHOW_ROI command, SHOW_NEXT reports the next continuous group of
channels that have the ROI flag set. The response is of the form:

$Dsssssnnnnnccc<CR>

where sssss represents an integer number that is the number of the first channel of the “next” group of

channels that all have their ROI bit set, and nnnnn represents an integer number that is the number of
channels in the group. If no more channels have their ROI bit set, SHOW_NEXT returns a first channel of 0 and
a number of channels of 0. The SHOW_ROI command is used to report the “first” group of channels that all

have their ROI bit set.
Example Responses:

$D0100000050078<CR>

Next ROI group starts at channel 1000 and is 50
channels long.

$D0215000150086<CR>

Next ROI group starts at channel 2150 and is 150
channels long.

$D0000000000072<CR>

No other ROI groups to report.

Summary of Contents for ORTEC EASY-NIM 928 Suite

Page 1: ...ORTEC EASY NIM 928 Suite High Performance Multi Function Nuclear Multichannel Buffer Counter Timer Rate Meter Hardware Manual Printed in U S A ORTEC Part No 932512 0114 Manual Revision A...

Page 2: ...ssion 865 483 2133 of the nature of the fault of the instrument being returned and of the model serial and revision Rev on rear panel numbers Failure to do so may cause unnecessary delays in getting t...

Page 3: ...GE SPECTROSCOPY 7 3 3 EASY NIM 928 COUNTER TIMER OPERATION 8 3 3 1 Starting the Easy NIM Front Panel Emulation Software 8 3 3 2 The Main Screen 8 3 3 3 Counter Timer Group Controls 10 3 3 4 The Instru...

Page 4: ...More Information 31 4 SPECIFICATIONS 31 4 1 MULTICHANNEL BUFFER 928 AND 928 MCB 31 4 1 1 Inputs and Outputs 32 4 1 2 Presets 32 4 2 COUNTER TIMER RATE METER ALL MODULES EXCEPT 928 MCB 33 4 2 1 Inputs...

Page 5: ...WARNING Indicates a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not...

Page 6: ...PROPERTY OF AMETEK INC ORTEC ORTEC EASY NIM 928 Suite 932512A 0114 vi...

Page 7: ...n be assigned to a variety of sources 1 1 FEATURES High performance multi function spectroscopy MCB combined with counter timer rate meters in a single width NIM module Group counters and timers for s...

Page 8: ...s and supplemental Windows applications programming interfaces that can be called from C C or Visual Basic For programmers using ActiveX controls all the functionality can be accessed more convenientl...

Page 9: ...e and disable the drivers for your ORTEC MCB s and share ORTEC MCBs across a network It also points you to information on selecting the proper network protocol for legacy MCBs At the end of installati...

Page 10: ...2 5 CONNECTING TO AND DISCONNECTING FROM THE PC The USB connection allows you to connect 928s to and disconnect them from a USB port without shutting down the computer or USB hub Note that if MAESTRO...

Page 11: ...ts positive unipolar positive gated integrator or positive leading bipolar analog pulses in the dynamic range from 0 to 10 V 12 V maximum ADC GATE Front panel optional TTL input Software selectable Co...

Page 12: ...Figure 3 Model 928 Full Counter Overflow n Pulse option 1 s wide pulse with rising edge synchronized to the rollover of a selected counter n 1 2 3 or 4 Level option On overflow output remains high unt...

Page 13: ...the selected counter input Software Control Hardware command SET_OUTPUT n x where n is the output and x is 0 or 1 3 2 CABLING FOR HPGE SPECTROSCOPY Figure 4 shows the standard cabling of a 928 or 928...

Page 14: ...en 3 3 2 The Main Screen Figure 6 shows the Easy NIM main screen Note that when you select a 928 COUNT 2 from the MCB list at the bottom of the screen only 2 counters and 1 timer are displayed Figure...

Page 15: ...t to Software Individual Reset the Reset button on this screen becomes active allowing you to manually zero the counter When a counter is enabled it remains active until disabled Data acquisition will...

Page 16: ...and reset options on the Instrument Settings screen They allow you to gang one or more counters and or timers together then enable disable and reset them simultaneously You can add additional counters...

Page 17: ...counter counts up to overflow 4294967295 counts then automatically resets to zero and continues to count up Fifo Click this button to display the values stored in the FIFO first in first out buffer s...

Page 18: ...PROPERTY OF AMETEK INC ORTEC ORTEC EASY NIM 928 Suite 932512A 0114 12 Figure 12 Counter Setting Lists Figure 13 FIFO Contents Figure 14 FIFO Contents Saved as Text File...

Page 19: ...a signal from any of the six front panel inputs or 4 simultaneously reset a group of timers or counters and timers with the Group 1 or 2 reset signal Enable input Use the Software Individual Enable t...

Page 20: ...en count rate exceeds this value the CRM gauge readout on the main screen flashes red In addition if the Play sound when alarm triggered checkbox is marked on the setup screen an audio alarm sounds If...

Page 21: ...Counter 1 The Gate inputs for both counters are set to Timer 1 Interval both are designated to reset on the Group 1 Reset and both are set to respond to the Software Group 1 Enable In this example Au...

Page 22: ...PROPERTY OF AMETEK INC ORTEC ORTEC EASY NIM 928 Suite 932512A 0114 16 Figure 20 Creating a Counter Timer Group Figure 21 Results for Group 1 Data Acquisition...

Page 23: ...matically In this continuous mode data collection continues until you click the timer s Disable button on the main screen Figure 22 shows the physical connections between the two 928s and from the eig...

Page 24: ...nd 928 and resets them when the Group 1 Reset signal is sent Each of the four Counter Inputs is configured for the type of external signal on its respective BNC Figure 24 shows the upper part of the I...

Page 25: ...riod acquisition stops until manually restarted In order to use the first 928 to control the counters on a second 928 the two MCBs must be cabled together as shown in Figure 26 For clarity this illust...

Page 26: ...screen Figure 27 Note that the timers and outputs in this illustration are faded for clarity they are not faded onscreen The four Counters are set to their respective counter inputs and each Counter I...

Page 27: ...le reset buttons on the main screen The Counter 2 source is the Counter 2 input and its Counter Input is configured for the type of signal on that BNC The Counter 2 and Timer 2 Gate input fields are b...

Page 28: ...utput 1 2 assignments are necessary The Enable input is set to MCB Active the Reset input is set to MCB Clear Command and we choose not to use Auto reset The Counter input is set to ROI This task does...

Page 29: ...m of the dialog Figure 28 ADC Tab Gate This control allows you to select a a positive TTL logic gating function With this function Off no gating is performed that is all detector signals are processed...

Page 30: ...s level establishes a lower level cutoff by channel number for ADC conversions The Upper Level Discriminator sets the level of the highest amplitude pulse that will be stored This level establishes an...

Page 31: ...DA presets are shown on a separate tab Figure 31 Presets Tab The presets can only be set when the 928 is not acquiring data You can use any or all of the presets at one time To disable a preset enter...

Page 32: ...or this MCB marked with an ROI reaches this value This has no function if no ROIs are marked in the MCB The Uncertainty preset stops acquisition when the statistical or counting uncertainty of a user...

Page 33: ...preset enter the preset value in the MDA or Correction field select the Nuclide and Energy enter the desired values for coefficients a b and c then click Add New To edit an existing preset click to h...

Page 34: ...nds the counting time to make up for the lost counts it will no longer be counting the same source as when the losses occurred As a result the number of counts in the peak will not be correct When a s...

Page 35: ...M_CORR the two spectra stored are the LTC spectrum and the ZDT spectrum corrected for the dead time losses real time only Unfortunately in the analysis of the ZDT spectrum the uncertainty of the measu...

Page 36: ...the time required to acquire the spectrum the standard mode ZDT Off is the preferred mode only the uncorrected spectrum is collected and stored in the spectrum file But if the counting rate varies sig...

Page 37: ...nt counting rate Because the error spectrum is not collected in NORM_CORR mode the standard deviation in the ZDT counts cannot be calculated if the counting rate is varying ISOTOPIC provides some prot...

Page 38: ...range from 0 to 10 V 12 V maximum semi Gaussian shaped time constants from 0 25 to 30 s gated integrator shaped time constants from 3 to 30 s or delay line shaped with widths 0 25 s Zin 1 k dc couple...

Page 39: ...pose TTL outputs Low is indicated by 0 4 V High is indicated by 2 4 V The outputs can drive 50 The outputs can be configured under software control to one of the following functions Timer Interval Out...

Page 40: ...ndividual Enable from software control The Group 1 or Group 2 enable from software control Overflow Output provides a rising edge at the moment a counter overflows The pulse can be routed to either of...

Page 41: ...the number of repetitions of the currently selected Time base FIFO Readout Control Timer FIFOs function the same as the counter FIFOs see above 4 3 ELECTRICAL AND MECHANICAL High Speed USB Standard co...

Page 42: ...s available 1 8 Sample mode functions available 0 9 Digital Offset e g 920 0 10 Software selectable analog offset 0 11 HV power supply 0 12 Enhanced HV SET_HV SET SHOW_HV_POL SHOW_HV_ACT 0 13 Software...

Page 43: ...0 37 MCS mode positive TTL input available 0 38 MCS mode fast negative NIM input available 0 39 MCS mode discriminator input available 0 40 Software switchable MCS mode discriminator edge 0 41 Softwa...

Page 44: ...0 SET_SHAP 1 0 62 Explorable gain ranges LIST VERI_GAIN_FINE _COAR _CONV 0 63 Explorable gain ranges LIST VERI_GAIN_FINE _COAR _CONV Beginning of 3rd word 0 64 External dwell support ENA DIS_DWELL_EXT...

Page 45: ...UM_DET SHOW_HV_RECO 0 90 Nuclide report SET SHOW_NUCL_COEF SET SHOW_ROI_NUCL 0 91 Interactive Display Features Such as Nuclide Report 0 92 Advanced Stored Spectra SH_SPEC_COUNT SET SHOW_SPEC_ID MOVE 1...

Page 46: ...page 0 108 PZ and flattop tilt not user changeable from property page 0 109 LFR not user changeable from property page requires Bit 96 0 110 Portal Monitor style List Mode Synch is available 0 111 DS...

Page 47: ...ive time or conversion gain The command SET_WINDOW 0 8192 has two parameters 0 and 8192 which set the window of interest to start at channel 0 and continue for 8192 channels Some parameters are option...

Page 48: ...numbers 128 No sample data available 5 3 DOLLAR RESPONSE RECORDS SHOW and STEP commands respond with a dollar response record The valid dollar response records for each command are discussed in the co...

Page 49: ...rting with a command prototype line Uppercase letters numeric digits blank space and special symbols such as the underscore _ and comma 13 in the prototype line are literal text to be sent to the MCB...

Page 50: ...overflow preset Channels that receive a count when they contain 2147483647 counts the maximum number of counts will stop the acquisition if the overflow preset is enabled The channel that caused the p...

Page 51: ...84 SET_GAIN_CONVERSION 512 Conversion gain set to 512 channels SET_GAIN_CONVERSION 1024 Conversion gain set to 1024 channels SET_GAIN_CONVERSION 2048 Conversion gain set to 2048 channels SET_GAIN_CONV...

Page 52: ...cquisition terminates SET_MDA_PRESET MDAPreset PeakNumber Sets the MDAPreset PeakNumber value as shown in the equation in the SET_MDA_COEF description above MDAPreset is usually the product of the des...

Page 53: ...e preset is complete and the acquisition is stopped Setting a true time preset to 0 ticks disables the preset See also CLEAR_PRESETS and SHOW_TRUE_PRESET SET_ULD value This sets the upper level discri...

Page 54: ...A 1 bit in the mask indicates that the corresponding segment is counting Responses C00000087 CR Neither input is acquiring data C00001088 CR Only input 1 is acquiring data C00002089 CR Only input 2 is...

Page 55: ...294967295 SHOW_LIVE Reports the contents of the live time counter in units of 20 ms 50 ticks s See also CLEAR_COUNTERS and SET_LIVE Responses G0000000000075 CR Live time reported as 0 ticks G000000000...

Page 56: ...for the state of health parameter for num SHOW_MONI_MAX Returns with the number of state of health values available for viewing SHOW_MONI_VALUE num Returns with an ASCII representation of the value f...

Page 57: ...if more that one channel contains the largest number of counts Channel 16383 is the highest numbered channel in any device Responses C00000087 CR Maximum count was found in channel 0 or no ROI channe...

Page 58: ...928 SHOW_SNUM Responds with a F record indicating the serial number of the EASYNIM 928 Response F100 Serial number 100 SHOW_STATUS Returns system status information in the following format Mlllllllll...

Page 59: ...s a C record SHOW_UNCERTAINTY Returns the current value of the uncertainty for the peak in the uncertainty preset See also SET_UNCERTAINTY Responses UNCE 0000000000008 5 Uncertainty of the peak is 8 5...

Page 60: ...any value from 0 to 65535 but is ignored 5 5 COUNTER AND TIMER COMMANDS ALL MODULES EXCEPT 928 MCB These commands control the counter timer functions in all but the 928 MCB For the 928 COUNT2 counters...

Page 61: ...ways on 1 Gate is connected to Input 1 BNC 2 Gate is connected to Input 2 BNC 3 Gate is connected to Input 3 BNC 4 Gate is connected to Input 4 BNC 5 Gate is connected to Input 5 BNC 6 Gate is connect...

Page 62: ...on status mask Read only Can only be used in SHOW_COUNTER_CONFIG Bit mapped flags indicate whether counter or timer has started counting Bit 0 Counter 1 Bit 1 Counter 2 Bit 2 Counter 3 Bit 3 Counter 4...

Page 63: ...nized to MCB acquisitions 8 Group 1 Reset 9 Group 2 Reset 2 Enable Value 0 MCB Active 1 Software Individual Enable 2 Software Group 1 Enable 3 Software Group 2 Enable 3 Timebase Value 0 Time Base is c...

Page 64: ...sy High when DAC state machine is busy transferring data to the hardware 0 AcquisitionActive High when acquisition is active SHOW_COUNTER_VALUE Returns a string with format COUNTER_VAL n1 n2 n3 n4 whe...

Page 65: ...932512A 0114 ORTEC EASY NIM 928 Suite 59 STOP_COUNTER n1 n1 is the counter index from 1 to 4 STOP_GROUP n1 n1 is the group index from 1 to 2 STOP_TIMER n1 n1 is the timer index from 1 to 2...

Page 66: ...Custom software applications 2 35 hardware commands and responses 41 Easy NIM software group control 10 installation 3 main screen 8 saving and recalling settings 14 setup Instrument Settings screen...

Reviews: