Maintenance | 4-15
Replacing Probe Heater Parts – Extended
Maintenance
This section describes Extended (Advanced) Maintenance, which is
required only in the event of failure of one of the Heater components
(Over-Temperature Switch, Heater Cartridge, or RTD). Replace the faulty
component(s) as required.
Carefully read and understand all of the instructions before begin-
ning disassembly and reassembly. The instructions are described in
a sequence that will ensure the proper removal and replacement of
all parts that make up the entire Heater Assembly. The instructions
also include special Notes, Cautions, and Warnings to consider while
working on the Heater Assembly.
IMPORTANT
If the Heater Assembly cover is removed for the purpose of replacing
internal electrical parts or wires,
always replace the internal o-ring
(P/N 300-8721).
Be sure to have the required parts (Over-Temperature Switch, Heater
Cartridge, and/or RTD) on-site prior to disassembling the Probe
Heater.
ORDER THE O-RING AT THE SAME TIME YOU
ORDER ALL OTHER PARTS.
While the Heater Assembly is removed for maintenance, take extreme
care to avoid scratching or damaging the joining surfaces (flamepath)
of the Heater Assembly and its cover, and the joining surfaces (heat
transfer) of the Heater Assembly and the HAG Probe. Also, take are to
avoid damaging the o-ring groove.
Gently clean these areas with a non-abrasive cloth.
If the Thermal Insulation Tube is removed, take extreme care to avoid
damaging its threads and the threads in the Heater Block, and the
Junction Box and its cover. After removing these parts, cover the
threaded ends to avoid damaging them. Clean, defect-free threads are
essential to ensure a flame-proof connection.
NOTE
NOTE
!
WARNING
!
CAUTION
See Chapter 5 for
extended maintenance spare
parts ordering numbers.
Summary of Contents for ASR900
Page 16: ...1 4 Heated Acid Gas Sample Probe Large Diameter Figure 1 2 HAG Sample Probe assembly drawing...
Page 25: ...Specifications 2 5...
Page 26: ...2 6 Heated Acid Gas Sample Probe Large Diameter...
Page 27: ...Specifications 2 7...
Page 28: ...2 8 Heated Acid Gas Sample Probe Large Diameter...
Page 29: ...Specifications 2 9...
Page 52: ...3 22 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Page 59: ...Maintenance 4 7 Figure 4 1 Large Diameter Heated Acid Gas Sample Probe Assembly...
Page 80: ...4 28 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Page 86: ...6 2 Heated Acid Gas Sample Probe Large Diameter Heated Acid Gas Sample Probe Assembly 100 2196...
Page 89: ...Drawings 6 5 Fiber Filter Assembly 100 2194...
Page 90: ...6 6 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Page 92: ...S 2 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...