3-18 | Heated Acid Gas Sample Probe (Large Diameter)
Wiring Sample Probe Heaters Controlled by Standalone
Temperature Controller (Retrofits)
A separate, external Temperature Controller device must be installed and
connected to provide signals and power to the HAG Probe Heater. This
device can be supplied by AMETEK or by the customer/end-user.
The Temperature Controller must be able to accept a European Curve
three-wire Pt100 RTD (0.00385 alpha coefficient), provide PID control, and
must also meet or exceed the following:
Temperature Range
40 °C to 120 °C, adjustable
Electrical Requirements
Must be able to switch a 150 W load at either 120 VAC or 240 VAC.
To wire the Probe Heater Assembly to the standalone Temperature
Controller:
1. Install the standalone Temperature Controller in a suitable location, as
close as possible to the HAG Sample Probe, to ensure the RTD in the
Sample Probe operates efficiently.
Hazardous Locations
The Temperature Controller must be located in a remote safe area, or
in a purged shelter/enclosure.
All electrical connections, materials, and methods (plus all safety poli-
cies and procedures) must be made in compliance with local wiring
regulations and electrical code for the hazardous area, and be approved
by the Owner Company.
2. The signal and power conductors can be run in separate or common
conduit (
North American/General Purpose
installations
) or cable
(
European/Hazardous Location
installations
). Terminal Box entries
are ½
"
NPT.
For wiring gauge (AWG) and other requirements, refer to draw-
ings/documentation shipped with the HAG Probe, analyzer, and/or
Temperature Controller.
NOTE
!
CAUTION
Summary of Contents for ASR900
Page 16: ...1 4 Heated Acid Gas Sample Probe Large Diameter Figure 1 2 HAG Sample Probe assembly drawing...
Page 25: ...Specifications 2 5...
Page 26: ...2 6 Heated Acid Gas Sample Probe Large Diameter...
Page 27: ...Specifications 2 7...
Page 28: ...2 8 Heated Acid Gas Sample Probe Large Diameter...
Page 29: ...Specifications 2 9...
Page 52: ...3 22 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Page 59: ...Maintenance 4 7 Figure 4 1 Large Diameter Heated Acid Gas Sample Probe Assembly...
Page 80: ...4 28 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Page 86: ...6 2 Heated Acid Gas Sample Probe Large Diameter Heated Acid Gas Sample Probe Assembly 100 2196...
Page 89: ...Drawings 6 5 Fiber Filter Assembly 100 2194...
Page 90: ...6 6 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Page 92: ...S 2 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...