Maintenance | 4-3
Frequency
Task
Other
Extended Maintenance:
Probe Heater
The Heater Assembly temperature zone circuitry does not require
regular maintenance. However, if problems arise with the Heater
(Probe-related alarms are active), determine the origin of the problem
and immediately replace the faulty part(s). These parts can include the
Over-Temperature Switch, Heater Cartridge, and RTD.
Refer to the “Maintenance and Troubleshooting” chapter (see section
“Host Controller Board Error Conditions and Corrective Action”) in the
analyzer manual for information on specific alarms. See “Replacing
Parts in the Probe Heater Assembly” in this chapter.
IMPORTANT
Be sure to have on-site the proper replacement parts prior to
disassembling the Heater Assembly. Refer to the “Service and
Parts” chapter for the part numbers.
Other
Flamepath Gap
During each Probe Heater Assembly maintenance, use a feeler gauge
to check the flamepath gap between the Heater Body and the Heater
Cover (see Figure 4-4 for location). The gap must not exceed 0.04 mm;
if the gap exceeds this value, contact AMETEK for advice. See
“Examining and Caring For the Flamepaths” in this chapter for more
information.
Summary of Contents for ASR900
Page 16: ...1 4 Heated Acid Gas Sample Probe Large Diameter Figure 1 2 HAG Sample Probe assembly drawing...
Page 25: ...Specifications 2 5...
Page 26: ...2 6 Heated Acid Gas Sample Probe Large Diameter...
Page 27: ...Specifications 2 7...
Page 28: ...2 8 Heated Acid Gas Sample Probe Large Diameter...
Page 29: ...Specifications 2 9...
Page 52: ...3 22 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Page 59: ...Maintenance 4 7 Figure 4 1 Large Diameter Heated Acid Gas Sample Probe Assembly...
Page 80: ...4 28 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Page 86: ...6 2 Heated Acid Gas Sample Probe Large Diameter Heated Acid Gas Sample Probe Assembly 100 2196...
Page 89: ...Drawings 6 5 Fiber Filter Assembly 100 2194...
Page 90: ...6 6 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Page 92: ...S 2 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...