4-2 | 3050-TE Moisture Analyzer
General Tab V2.0
Use the
General
tab to view the current configuration and define parameters for
the analyzer and for
PC communications
(Figure 4-1).
After communication with analyzer is established, any changes to the
analyzer communications parameters must be made using the Device
Communications tab. These changes must be made before you make any
changes to the computer serial port settings.
PC Communications
Click the
Setup
button to configure
PC Communication
s. The
Serial Port Com-
munication
screen opens (Figure 4-2).
Communication Protocol
Select AMETEK Serial for inital PC communication setup. Once
communications with the analyzer are established, the Modbus
serial can be selected if desired.
When changing from AMETEK Serial to Modbus Serial, the analyzer
communications parameters must be changed before the computer serial
port settings.
NOTE
Figure 4-1.
General tab
Configurator software.
Figure 4-2.
PC Serial
Communications setup
screen.
2.2
NOTE
Summary of Contents for AMEVision 3050-TE
Page 9: ...Contents ix...
Page 16: ...1 6 3050 TE Moisture Analyzer This page left intentionally blank...
Page 20: ...2 4 3050 TE Moisture Analyzer This page intentionally left blank...
Page 25: ...Installation Start Up 3 5 Figure 3 3a Division 1 Installation Dimensions...
Page 26: ...3 6 3050 TE Moisture Analyzer Figure 3 3b Zone 1 Installation Dimensions...
Page 33: ...Installation Start Up 3 13 Figure 3 10 RS 485 Cables Multiple 3050 s...
Page 34: ...3 14 3050 TE Moisture Analyzer Figure 3 11 RS 485 to RS 232 Coversion for Host PC...
Page 51: ...PC Interface 4 15 Figure 4 11b Flow Adjust Screen...
Page 52: ...4 16 3050 TE Moisture Analyzer This page intentionally left blank...
Page 54: ...5 2 3050 TE Moisture Analyzer This page intentionally left blank...
Page 60: ...Replacement Parts Service 6 6 This page intentionally left blank...