A-14 | 3050-TE Moisture Analyzer
ID/STATUS INFORMATION
The MODBUS master can poll the analyzer periodically for status infor-
mation via MODBUS function 17 (11Hex). The returned information has
the following format:
1 byte Slave ID
= 50h for Model 3050 analyzer
1 byte Run Status
= FFh for Analyzer on Line
(invalid signal = 0)
= 00h for Analyzer off Line
(invalid signal = 1)
2 bytes Status Word
= System State which is register #8
The most significant byte comes first.
15 bytes Model Name
=Analyzer Model Name are registers
213-220
12 bytes Serial Number
=Analyzer Serial Number located in
registers
183-189
4 bytes Version Number
=S200 located in register #82
The byte count is 35 (23Hex).