A-10 | 3050-AMS Moisture Analyzer
Data
Capture
The data capture feature allows the user to collect and save analyzer data
displayed on the Monitor screen to an Excel compatible file.
On
Press the ON button to start data collection. Specify the file name in the
“Save As” dialog and press “Save” button. The file format is “.csv” which
is Excel compatible. All data displayed on the Monitor page will be stored
in the specified file.
Figure 3-10b. Monitor
tab - Save As.
Off
Data collection will terminate and the file will close when the Off button is
pressed or another configurator tab is selected.
Rate
The preferred data collection rate is “0.5” minutes and is set as the default.
A record is created every 30 seconds. The collection rate can be increased
to one minute or more.