7-14 | 3050 AMS Moisture Analyzer
NOTE
Pilot Valve Manifold
The Pilot Valve Manifold assembly (AMETEK Part #583011901) is found
on the side of the Power Supply PCB (Figure 7-7).
Removing the pilot valve manifold
1. Using a #2 Philips screwdriver, loosen the two captured screws that
hold the pilot valve manifold assembly to the sample manifold plate
(Figure 7-7 and Figure 7-11)
2. Disconnect the pilot valve receptacle plugs (Figure 7-11).
3. Use a 1/16” slotted screwdriver to remove the two (2) screws at the
side of the manifold (Figure 7-11).
The mounting screws for the two manifolds are different size.
4. Detach the tubing from the manifold by holding the tubing retaining
collar and pulling the tubing free. Make sure that you identify the tub-
ing for proper position when replacing.
Figure 7-11.
Model 3050-AMS pilot
valve manifold.
Receptacle Plug
Screws
(
Use Slotted
Screwdriver
)
Summary of Contents for 3050-AMS
Page 12: ...xii 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 22: ...2 4 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 25: ...Installation Start Up 3 3 Figures 3 1c Model 3050 AMS dimensions top and sides...
Page 42: ...3 20 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 64: ...5 10 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 66: ...6 2 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 84: ...7 18 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 96: ......