Serial Communications | 5-5
Write Data (H)
The data field of this frame contains the handle for a system variable. See
Figures 5-7a and 5-7b for a list of defined handles for system variables.
Example
:
Master:
>34H05123.40C[CR]
Slave:
A[CR]
Write Cell EEPROM (L)
Copy cell serial number and cell calibration coefficients into cell EEPROM.
Example
:
Master:
>34LB3[CR]
Slave:
A[CR]
Quit (Q)
Zero cycle is terminated immediately and normal operation resumed .
Example
:
Master:
>34QB8[CR]
Slave:
A[CR]
Zero (Z)
Initiates a zero cycle.
Example:
Master
:
>34Z??[CR] (
Note the use of wildcard in place of checksum)
Slave
:
A[CR]
Summary of Contents for 3050-AMS
Page 12: ...xii 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 22: ...2 4 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 25: ...Installation Start Up 3 3 Figures 3 1c Model 3050 AMS dimensions top and sides...
Page 42: ...3 20 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 64: ...5 10 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 66: ...6 2 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 84: ...7 18 3050 AMS Moisture Analyzer This page intentionally left blank...
Page 96: ......