7-4 | 3050 AMS Moisture Analyzer
Capillary
The capillary (Figure 7-4) is a part of the flow meter in the analyzer and
can easily be replaced when necessary.
Removing the capillary
1. Remove the top cover from the analyzer and remove the cover from
the oven assembly. Facing the front of the Model 3050-AMS analyzer,
the capillary is found on top of the stream switching manifold (Figure
7-4).
3. Disconnect the capillary and remove it by loosening both nuts and
gently pulling the tubing out of the stream switching manifold block.
Installing the capillary
1. To install the capillary, gently push the tubing ferrules on the capillary
into the tubing on the manifold block until flush and tighten the nuts
to secure.
2. Replace the oven lid and secure it with the two nuts. Replace the ana-
lyzer cover and secure it with the mounting screws.
Figure 7-4.
Capillary location - 3050-
AMS.
Capillary
Sensor
Proportinal
Solenoid
Valve
Proportinal
Solenoid
Valve
Manifold