5 - 11
User Guide
LMG MkII
Signal Processor
5.3.5 Cable lengths
The maximum cable lengths are, to a great extent, dependent on baud rate
and cable type. Generally a lower baud rate will give reliable communications
over a longer length of a given cable type.
RS485
9600 baud, 1000 m when using the cable recommended below,
with correct cable terminations.
Recommended 1000 m cable for RS485:
•
2 pairs individually screened;
~50 ohms characteristic impedance, <= 55 ohm/km;
<100 pf/m core to core
<200 pf/m core to screen