User Guide
5 - 8
Signal Processor
LMG MkII
5.2.3 RS485 examples
The RS485 communications is designed primarily for use by a computer
running a customised communications package.
To read the temperature from channel ‘A’ ...
Type at the terminal
[STX][addr]RATMP[ETX]
breakdown
STX Binary 0x02, start of text framing character
addr Binary address of processor to be interrogated
R
Read mode
A
Channel A
TMP Command to report the temperature
ETX Binary 0x03, end of text framing character
reply
[STX][addr]A1000C[ETX]
breakdown
STX Binary 0x02, start of text framing character
addr Binary address of replying processor
A
Channel ident of replying processor
1000C Temperature and units
ETX Binary 0x03, end of text framing character
To set the emissivity of channel ‘B’ to 0.5 ...
Type at the terminal
[STX][addr]SBEMS 50[ETX]
breakdown
STX Binary 0x02, start of text framing character
addr Binary address of processor to be interrogated
S
Set mode
B
Channel B
EMS Command to work on the emissivity value
50
New emissivity value
ETX Binary 0x03, end of text framing character
reply
Will reply with “[channel letter]*” if command fails
Summary of Contents for LMG MkII
Page 6: ...INTRODUCTION 1 ...
Page 10: ...INSTALLING THE PROCESSOR 2 ...
Page 14: ...User Guide 2 4 Signal Processor LMG MkII Fig 2 4 LMG MkII Electrical System Overview ...
Page 22: ...3 USING THE PROCESSOR ...
Page 44: ...TIME FUNCTION PROCESSING 4 ...
Page 50: ...SERIAL COMMUNICATIONS 5 ...
Page 62: ...Signal Processor LMG MkII Blank ...
Page 63: ...EXTERNAL I O MATHS FUNCTIONS 6 ...
Page 64: ...Signal Processor LMG MkII Screws A Fig 6 1 Location of screws on back panel Screws A ...
Page 88: ...MAINTENANCE 7 ...
Page 92: ...User Guide 7 4 Signal Processor LMG MkII Fig 7 3 USB Export Data option ...