5 - 5
User Guide
LMG MkII
Signal Processor
5.1.2 Maths Parameters (Common Parameters)
For a explanation of Maths Functions, see Section 6.
Name
Menu title text
Range
Increment
Default
Comments
Min
Max
Units Menu value text
MT1 Maths Function 1 Type
0
5
1
0
0 = Max
1 = Min
2 = Mean
3 = Diff
4 = Range
5 = BGComp
MC1 Maths function 1
Allocation
Add together:
ChA = 128
ChB = 64
ChC = 32
ChD = 16
MH1 Maths Function 1
Tmax
1
Dependent upon
thermometer
ML1 Maths Function 1
Tmin
1
Dependent upon
thermometer
MA1 Maths Function 1
Alarm 1 Level
1
Dependent upon
thermometer
MM1 Maths Function 1
Alarm 1 Mode
0
1
1
0
MO1 Maths Function 1
Output Mode
0
1
1
1
mA
0 = 0 to 20 mA
1 = 4 to 20 mA
ST1
Maths Function 1
Temperature
System output
value. Read only
SS1
Maths Function 1
Status
0
255
Bits in the Status
byte:
0 = Not set
1 = Set
2 = Alarm A2
3 = Error
4 = Over range
5 = Under range
6 = Units
7 = Alarm A1
Summary of Contents for LMG MkII
Page 6: ...INTRODUCTION 1 ...
Page 10: ...INSTALLING THE PROCESSOR 2 ...
Page 14: ...User Guide 2 4 Signal Processor LMG MkII Fig 2 4 LMG MkII Electrical System Overview ...
Page 22: ...3 USING THE PROCESSOR ...
Page 44: ...TIME FUNCTION PROCESSING 4 ...
Page 50: ...SERIAL COMMUNICATIONS 5 ...
Page 62: ...Signal Processor LMG MkII Blank ...
Page 63: ...EXTERNAL I O MATHS FUNCTIONS 6 ...
Page 64: ...Signal Processor LMG MkII Screws A Fig 6 1 Location of screws on back panel Screws A ...
Page 88: ...MAINTENANCE 7 ...
Page 92: ...User Guide 7 4 Signal Processor LMG MkII Fig 7 3 USB Export Data option ...