Troubleshooting Guide
9.1.1 Warning & alarm messages
Displayed Message or
notification
Reason
Checks
Proposed Actions
“Tank Low Level”
The internal water
break tank has
insufficient water
to run the unit.
1.
Make sure the
feedwater supply is
turned on and
flowing. Check
other taps to
confirm.
1.
If the feedwater supply is
not a problem and the unit
still does not run and this
message is frequently seen
contact your Healthcare
provider.
2.
If the water supply has
stopped contact your water
utility company.
“High Pressure”
The unit has
detected an
unsafe operating
pressure.
1.
There are no
checks to made,
simply switch the
unit off using the
switch at the rear
and turn off the
water supply.
1. DO NOT
try to run the unit
in this condition, contact your
Healthcare provider for
assistance.
“Low Pressure”
The unit has
detected
insufficient
pressure to
operate.
1.
Check that the
feedwater is still
flowing.
2.
Check that there
are no leaks.
3.
Check for any
other messages on
the display.
1.
If the checks do not show
any problems, press the
“START” button. After a few
minutes, if the pressure in
unit has returned to normal
the unit will run. If the
message returns then switch
the unit off, turn off the water
supply and contact your
Healthcare provider for
advice.
“Perm-Line Cell
Error”
The unit has
detected a fault
with the sensor
measuring the
water quality.
1.
There are no
checks to be made.
1.
For safety the unit will
have stopped running as the
quality of water being
produced cannot be
monitored accurately.
Contact your Healthcare
provider to resolve the
problem.