23
7.2
Heat disinfection
As a guide refer to
Section 8.1
for recommendations as to when to carry out
heat disinfection on the unit.
Warning:
Heat disinfection should only be carried out if you have been trained by
AmeriWater
or by your
Healthcare provider. For security a key is provided once you have been trained which will
enable you to activate the heat disinfection cycle.
DO NOT
attempt to clean the unit if you have
not been trained.
If you wish to clean the unit and carry heat disinfections yourself please contact
AmeriWater
or your Healthcare provider who will provide all the necessary instructional training, instruction
leaflet and security key to enable the heat disinfection to take place.
Failure to heat disinfect the unit or carry out recommended chemical cleans could result in a
deterioration of the water quality provided by the unit making it unsuitable for hemodialysis.
During heat disinfection of the unit there is a potential for water exceeding 176
0
F to be
discharged to drain, ensure the drain pipework is made of suitable material to tolerate this water
temperature
8.0 MAINTENANCE AND CHECKS
8.1
Maintenance
Your Centurion unit will be maintained and serviced regularly by your
Healthcare provider. The unit does not contain any serviceable parts that
require replacement on a regular basis but will require servicing at set intervals
in order to ensure the unit’s performance. Refer to the
Service & Maintenance
manual
for details
As part of the servicing program for the unit certain components will require
replacing. Your Healthcare provider will be responsible for replacing these
parts.
It is recommended that regular checks are carried out on the unit and its
performance to ensure safe and uninterrupted operation. Refer to table below
for details.