background image

3.4  Intended Use 

The AmeriWater 

Centurion 1500+

 Reverse Osmosis Systems are water treatment 

systems intended for use in hemodialysis applications.  They are designed to pre-treat 
and purify potable water for use in making dialysate for hemodialysis and to meet 
current AAMI and Federal (U.S.) standards.  The device is intended to be a component 
in a complete water purification system, and is not a complete water treatment system.  
It must be preceded by pre-treatment devices, and may need to be followed by post-
treatment devices as well to meet current AAMI and Federal (U.S.) standards.  The 
Centurion 1500+ Reverse Osmosis System is intended for use in a hospital, clinic, 
dialysis center, or for home care for single patient use.  The device includes an 
integrated heat sanitization process. 

The 

Centurion  1500

+

  unit  is  a 

Class  III  (Health  Canada)  /  Class  II  (USA)  Medical 

Device

  intended  for  use  in  hemodialysis  applications.   It  should  only  be  used  to  feed 

hemodialysis equipment that complies with the current IEC 60601-2-16 standard. 

The  unit  has  been  designed  to  pre-treat  and  purify  potable  water  for  use  in  the 
preparation  of  dialysate  solution  suitable  for  hemodialysis  and  related  therapies  in 
accordance with current 

AAMI/ANSI/ISO

 and Federal (US and Canadian) standards.  

The 

Centurion  1500

+

  unit  is  intended  as  a  stationary  device  for  indoor  use  only  in 

hospitals,  clinics,  dialysis  centres  or  for  home  care  for  single  patient  use  as  part  of  a 

Medical Electrical (ME) system

. The unit should not be stacked  on top off or placed 

directly adjacent to other electrical/electronic equipment. 

The 

Centurion 1500

+

 water purifier works on the principle of reverse osmosis and Ultra-

filtration  to  provide  purified  water  suitable  for  both  hemodialysis  and  Hemodiafiltration 
and has been designed for continuous operation.  

CAUTION:   

When used as a medical device, Federal law restricts this device to sale by or 
on the order of a physician per 21CFR 801.109(b)(1).

3.4.1    Operating staff 

Warning: 

The unit must only be operated and installed by persons who have been suitably trained and have studied 
and understood the instructions within this manual and who are familiar and confident with the operation 
of the unit. 

Maintenance  is  restricted  to  carrying  out  chemical  cleans  and  heat  disinfections,  once  approved  and 
trained to do so by your Healthcare provider. There are no operator serviceable parts to be replaced. 

It is essential that in the event of an emergency the location of the Circuit Breaker or other supply n mains 
isolation device protecting the equipment is known.  

If  at  any  time  you  are  unsure  about  the  electrical  safety  of  the  unit  or  have  reason  to  believe  that  it  is 
potentially unsafe to use you should switch the unit off, isolate it at the Circuit Breaker and seek advice 
from your Healthcare provider. 

The supply mains isolation switch is located at the rear of the unit ensure that it is accessible at all times 
and remains clear from obstruction. Ensure that the wall plug is accessible at all times as this can be used 
as a method of isolation. 

Summary of Contents for CENTURION 1500+

Page 1: ...CENTURION 1500 OPERATING MANUAL P N 98 2010 REV C...

Page 2: ......

Page 3: ...Revision C December 2017 CENTURION 1500 UF OPERATING MANUAL For L998374 5 DOCUMENT HISTORY Revision Date of Change A January 2013 B January 2016 C December 2017...

Page 4: ...ures and process of operation 12 4 5 Explanation of rear connections 13 5 0 OPERATING THE TOUCHSCREEN 14 5 1 Explanation of Buttons 14 5 1 1 Using the screen 14 5 1 2 MENU button 14 5 1 3 START button...

Page 5: ...6 Feed water requirements 37 11 1 7 Water services connections 37 11 1 8 Raw water break tank 38 11 1 9 Weights and dimensions 38 11 1 10 Main PCB battery 38 11 1 11 USB mass storage device 38 11 2 En...

Page 6: ...r a qualified technician to maintain and service the unit Details on how to install commission the Centurion 1500 can be found in the Installation Commissioning Guide Installation of the unit would al...

Page 7: ...ly contact your Healthcare provider or AmeriWater or refer to the relevant accompanying manuals and guides detailed in Section 1 0 PREFACE AmeriWater Tel No 800 535 5585 Or your local authorized Ameri...

Page 8: ...formation that can be used to help explain a command or action or give guidance 3 2 Explanation of labels There are a number of labels applied to the outside of the Centurion 1500 unit which identify...

Page 9: ...th this label during heat disinfection or take suitable measures to protect yourself from the heat Do Not Sit Under no circumstances should the unit be sat on or objects placed on the top sloping cove...

Page 10: ...ean This label identifies the position of the key When approved trained to do so always refer to the Chemical cleaning and heat disinfection instructions if unsure about when and how to use this key E...

Page 11: ...tion 11 1 1 The Refer to the manual symbol will direct your Healthcare provider installer to the Installation Commissioning guide This label provides details of all the water connections If more detai...

Page 12: ...iately take the necessary precautions if the spillage contains bodily fluids Refer to Section 8 3 Cleaning external surfaces when dealing with spillages If the unit develops a leak follow the procedur...

Page 13: ...as part of a Medical Electrical ME system The unit should not be stacked on top off or placed directly adjacent to other electrical electronic equipment The Centurion 1500 water purifier works on the...

Page 14: ...ot designed to take the weight of the unit Support the unit by holding the underside of the chassis while steadying the body When relocating the unit always ensure the unit has performed its full drai...

Page 15: ...ned and operated according to the instructions contained within this Operating manual and supporting documents AmeriWater Healthcare provider will not accept any responsibility for any damage or injur...

Page 16: ...t The unit is fitted with an internal pump that pressurises the water supply and forces it through a membrane which then separates out all of the impurities from the feedwater Purified water termed pe...

Page 17: ...tank and boosts the water pressure to the RO membrane There is a pressure sensor situated immediately after the pump which will display a warning message on the touch screen should the maximum system...

Page 18: ...13 4 5 Explanation of rear connections Key Switch Alarm Output USB port On Off switch Fuses Power lead socket Acid cleaner inlet Water In Water Out Water Return Drain Cooling Fan...

Page 19: ...ion simply touch the screen on the appropriate button once with your finger Note Do Not use sharp or pointed implements such as pens pencils etc to operate the screen as this will damage the sensitive...

Page 20: ...tch to the I position When powered the unit s HMI screen will illuminate and display both text and a visual mimic of the units operation The touch screen will initially display Intialization for a few...

Page 21: ...e STOP button twice from the processing screen will stop the unit immediately terminating the flush Refer to Section 3 5 Warning The supply mains isolation switch is located at the rear of the unit en...

Page 22: ...measured in microsiemens cm or S cm Temperature This tells you the temperature of the water being produced and is displayed in 0 F Pump Pressure The unit contains a pump to process the water and this...

Page 23: ...ch are only provided for use by your Healthcare provider The various sub menu options can be selected by pressing the buttons followed by ENTER 6 1 1 Set Time and Date From the USER MENU and by using...

Page 24: ...p stating Done and after a few seconds the screen will revert back to the Time Date Menu To select Set Date from the menu use the buttons and press ENTER The date will be displayed and the cursor will...

Page 25: ...ecomes due 6 1 3 Product Information This screen shows the product details and software version numbers and is for information only 12 04 12 6 MONTH INTERVAL Pre Filter Date RESET BACK Product Informa...

Page 26: ...ettings Menus are only accessible for use by trained and authorized personnel If any data held within these menus is altered the performance safety of the unit and or water quality may be seriously af...

Page 27: ...to carry out cleaning disinfection of the unit When approved to do so ensure that the procedures detailed within this document are followed Warning Chemical cleaning of the unit MUST only be carried b...

Page 28: ...lt in a deterioration of the water quality provided by the unit making it unsuitable for hemodialysis During heat disinfection of the unit there is a potential for water exceeding 1760F to be discharg...

Page 29: ...ed water output Min 0 40 USgals min 500F Monthly Check to be carried out by Healthcare provider Hot water disinfection Min temp 1760F Min hold time 10 mins Once per week min Record of last heat disinf...

Page 30: ...r leaks in or around the unit Verify the conductivity on the Centurion display is below the alarm set point Note Do Not start dialysing if the unit appears to have a fault or the water quality is in d...

Page 31: ...0F On starting the unit up after standing run for 10 minutes disconnected from the dialysis machine and carryout a heat disinfection Your Healthcare provider will advise you on the best method of main...

Page 32: ...follow the instructions in Section 3 5 to shut the unit down quickly Once the unit has been safely shut down and the water supply turned off make a note of any messages that were displayed on the scre...

Page 33: ...e water supply 1 DO NOT try to run the unit in this condition contact your Healthcare provider for assistance Low Pressure The unit has detected insufficient pressure to operate 1 Check that the feedw...

Page 34: ...d not tilted over 1 To prevent the possibility of water leakage due to the internal tank overflowing for safety it will have stopped running 2 No other warning messages are displayed and the unit appe...

Page 35: ...ngineer has switched the unit off many times to repair a fault 1 Ensure that the mains electrical lead is secure and fixed 2 Check the condition of your circuit breaker and any other household applian...

Page 36: ...our Healthcare provider for advice Poor Water Quality The unit has detected that the quality of the permeate water is above 200 S cm Refer to Service Due message previously detailed Note Should the pe...

Page 37: ...entilation fan is not obstructed or the unit covered by anything and placed in a well ventilated area at ambient temperature away from any heat sources 1 Advise your Healthcare Provider immediately do...

Page 38: ...r trained qualified electrical engineer that is familiar with this equipment There is insufficient flow to run the dialysis machine 1 A fault has occurred with the unit that has either stopped the uni...

Page 39: ...ossible causes Checks Proposed actions Bacterial count and or Endotoxin levels in dialysis water exceed recommended guidelines 1 Fault with Heat disinfection cycle 2 Period between Heat disinfections...

Page 40: ...efer to section 8 2 Pre dialysis checks to assess performance of RO membrane 3 Repeat samples to confirm levels 4 Check pre treatment is functioning correctly 5 Refer to Section 8 1 1 for detail of ch...

Page 41: ...ective earth For permanent installations the mains supply must be provided with a Branch Circuit Breaker refer to Section 4 6 2 of the Installation Commissioning Guide for details of rating and specif...

Page 42: ...utput water quality 0 29 0 4 60 Will meet the requirements of current AAMI ANSI ISO 13959 standard for water for hemodialysis and related therapies 11 1 6 Feed water requirements Pre filtration Filter...

Page 43: ...pe Lithium CR2032 11 1 11USB mass storage device Specification FAT 16 formatted USB memory stick Memory size Must be less than 2GB 11 2 Environmental data Parameter Normal Operation Storage Transport...

Page 44: ...ssions CISPR 11 Group 1 The Centurion 1500 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equip...

Page 45: ...l commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth A A Mains power quality should be that of a typical commercial or hospital environment Voltage dips...

Page 46: ...ers m Field strength from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance lever in each frequency range b Interference may occur in the vicin...

Page 47: ...rs m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufactur...

Page 48: ...he device bears the following UL mark CLASSIFIED BY UNDERWRITERS LABORATORIES INC WITH RESPECT TO ELECTRIC SHOCK FIRE MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH ANSI AAMI ES60601 1 2005 3rd ed and CAN...

Page 49: ...thout deforming it or restricting flow Push the tube into the connector until the tube stops Ensure the tube is pushed in as far as the tube stop pull on the tube slightly to check that it is secured...

Page 50: ...nent from the unit must be in accordance with local requirements in your province or state for the disposal of electrical waste E Waste Your Healthcare provider will be responsible for the disposal of...

Page 51: ...46 12 3 Flow schematic...

Page 52: ...47 12 4 Certificate of conformity...

Reviews: