11.6 Separation distances for RF devices and Centurion
Recommended separation distances between portable and mobile RF communications equipment
and the Centurion 1500
+
.
The Centurion 1500
+
is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. AmeriWater or your Healthcare Provider will help prevent electromagnetic
interference by calculating and maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Centurion unit as recommended below, according to
the maximum output power of the communication equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d =
[1.17] √
P
80 MHz to 800 MHz
d =
[1.17] √
P
800 MHz to 2.5 GHz
d =
[2.33] √
P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
11.7 Classification and standards applied
Electrical
Class I Equipment (see Note1)
Overvoltage category – Fixed installation
Category II
Overvoltage category – Permanently installed
Category III
Pollution degree classification
Degree 2
IP Classification
IP21
Medical class - USA
Class II
Medical class - Canada
Class III
Radio Performance - USA
To FCC Part 18 (industrial, Scientific
and Medical)
Radio Performance - Canada
To ICES-001 (Industrial, Scientific and
Medical (ISM) Radio Frequency
Generators.
Designed in general accordance with the requirements of BS EN 60601-1-2 :2007
Medical Electrical Equipment – Part 1-2: general requirements for basic safety and essential
performance
Warning: Note 1,
To avoid risk of electric shock, this equipment must be connected to a supply
main with protective earth.