“Tilt”
The unit has been
moved or knocked
whilst running to
such an extent
that the internal tilt
switch has been
activated.
1.
Check that the
unit is upright and
standing on a firm
flat surface.
2.
Make sure that
there is nothing
located near the
unit that might
knock it .
3.
Check that the
unit is not placed
next to a door or
could be knocked
over by children or
pets.
1.
At the front underside of
the unit is a black drain plug,
unscrew the plug and let any
water drain from the unit,
then with the unit upright and
all potential hazards
removed, switch the unit off
and on using the on/off
power switch at the back of
the unit, the unit should now
run. If not and the message
remains call your Healthcare
provider for advice.
“Reset Error”
The unit has
switched on and
off more than 3
times in one
minute.
1.
Make sure the
mains electrical
lead is secure at
the connection to
the unit at the wall
socket.
2.
Check that
you’re household
circuit breaker has
not tripped out or
has a fault.
3.
Are you
experiencing power
cuts to your
property?
4
. Is the unit being
serviced and the
Engineer has
switched the unit off
many times to
repair a fault.
1.
Ensure that the mains
electrical lead is secure and
fixed.
2.
Check the condition of
your circuit breaker and any
other household appliances
being used. This may have
to be carried out you’re your
Healthcare provider.
3.
Check with your local
power company if they are
experiencing any power
losses in your area.
4.
If the problem persist and
none of the above are the
cause then contact your
Healthcare provider for
assistance.
“Feed-Line Cell
Error”
The unit has
detected a fault
with the line cell
used to measure
the quality of the
incoming
feedwater supply.
1.
Check that there
are no other
displayed alarm
messages, if there
are make a note of
them.
1.
Contact your Healthcare
provider and give them the
details of all messages
displayed in order for them
to diagnose the problem and
advise you if the unit can be
operated with this fault.
“Clean Due”
The time period
set by your
Healthcare
provider for the
next chemical
clean has expired
and the unit now
requires a
chemical clean.
No checks
required.
If you are trained and
approved to do so carryout a
chemical clean on the unit at
the next convenient time. If
you are not authorized to
clean the unit then contact
your Healthcare provider to
clean the unit.