69
Figure 10-2: Sensor Service Menu Flow Diagram
ADAPTICal
Window
Override
Reset
Exit
From Main
Service Menu
Background
+0.00
Select
ADAPTICal
Use
Up/Down
buttons to
navigate list and
press
Enter
Button
to select Sub-
Menu, Reset to
reset the Sensor
Module or Exit to
return to the Main
Service Menu
Window
+100
Select
Window
Enter the Window Filter value
one digit at a time (including
the sign) using the
Up/Down
Buttons. Press Enter to
advance to next digit. Press
Enter
button
while last digit is
highlighted will save value.
Save?
Yes
Use
Up/Down
buttons to toggle
between available options (Yes /
No) and press
Enter
Button to
select desired response.
Sim Enable
Sim Gas
Exit
Select
Override
See Override Sub-
Menu Flow
Diagram
Span Level
+40.00
ADAPTICal
“
Current
State
”
Abort
Enter the Background Gas level one digit at a
time (including the sign) using the
Up/Down
Buttons. Press Enter to advance to next
digit. Press
Enter
button
while last digit is
highlighted will save value.
Enter the Span Gas level one digit at a time
(including the sign) using the
Up/Down
Buttons. Press Enter to advance to next
digit. Press
Enter
button
while last digit is
highlighted will save value.
ADAPTICal
“
Cal Results
”
OK
Once calibration is complere the results will
be displayed as
“
Cal Passed
”
or
“
Cal Failed
”
with OK highlighted. Press
Enter
button
to
return to the Sensor Service Menu screen.
Calibration process has started and the menu will
display following states during process
•
Buffering
–
Unit is preparing for calibration
•
Find Zero
–
Unit is determining Zero Gas
value
•
Apply Gas
–
User to start applying
calibration gas
•
Span Level
–
Calibration is underway
–