AMC-UTx-M-xx-400 Series Digital Transmitter
35
Table 5-2: Service Menu Options
Menu Item
Description
<Sensor #> : <Gas Label>
Each sensor is included in the service menu indicated by its
number and gas label (Ex: S2 : NO2)
UTx
Access the settings for the transmitter module through the ‘UTx’
menu item
Exit
Select this menu item to exit the service menu
5.2.6 UTx-M Menu System
–Service Menu
Press
DOWN
repeatedly until the end is reach and Exit is highlighted as shown below.
With ‘Exit’ highlighted (as above), press
ENTER
to exit the service mode and return back to the
main info screen as shown below.
The service menus remains unlocked for a programmed timeout period (currently set to 5
minutes). Within that period, the service menus can be re-enabled by long pressing the ENTER
button. If the timeout period is allowed to expired, the password entry screen will appear prior to
accessing the service menus. Also, if the service menu is active when a timeout occurs, the
service mode is disabled, and the display will go blank (i.e. power saving mode). Lastly, the
service menu timeout is reset with each button press while the service menus are active.