NONE
- If this option is selected then remote input 1 will have no function.
•
P.HoId
- peak hold. The display will show the peak value (highest positive value) only whilst the remote
input terminals are short circuited i.e. the display value can rise but not fall whilst the input terminals are
short circuited. The message
P.HLd
will appear briefly every 8 seconds whilst the input terminals are short
circuited to indicate that the peak hold function is active.
•
d.HoId
- display hold. The display value will be held whilst the remote input terminals are short circuited.
The message
d.HLd
will appear briefly every 8 seconds whilst the input terminals are short circuited to
indicate that the display hold function is active.
•
P.Hi
- peak memory. The peak value stored in memory will be displayed if the remote input terminals are
short circuited, if the short circuit is momentary then the display will return to normal measurement after 20
seconds. If the short circuit is held for 2 to 3 seconds or the power is removed from the instrument then the
memory will be reset.
•
P.Lo
- valley memory. The minimum value stored in memory will be displayed. Otherwise operates in the
same manner as the
P.Hi
function described above.
•
HiLo
- toggle between
Hi
and
Lo
displays. This function allows the remote input to be used to toggle
between peak and valley memory displays. The first operation of the remote input will cause the peak
memory value to be displayed, the next operation will give a valley memory display.
P.Hi
or
P.Lo
will
flash before each display to give an indication of display type.
•
AI.Ac
- alarm acknowledge. Allows the remote input to be used to acknowledge a latching alarm. See the
Ltch
function.
•
ACCSS
- remote input access. Allows the remote input to be used for access control purposes. See the
R.INP LEUL
function.
•
P.but
- allows the selected remote input to act as the
P
pushbutton.
•
F.but
- allows the selected remote input to act as the
F
pushbutton.
•
U.but
- allows the selected remote input to act as the
^
pushbutton.
•
d.but
- allows the selected remote input to act as the
v
pushbutton.
•
duII
- display brightness control. The remote input can be used to change the display brightness. When
this mode is selected the display brightness can be switched, via the remote input terminals, between the
brightness level set at the function and the brightness level set at the function.
•
b.Inc
- display brightness increment. Allows the selected remote input to be used to increment the display
brightness. If no other remote input is set to
b.dEc
then once maximum brightness is reached the brightness
will start again from lowest brightness and increment. If auto brightness is turned on then when powered up
the display will go to its auto brightness level and can be adjusted from there. . If auto brightness is turned
off then the display will remember its last brightness setting on power up.
•
b.dEc
- display brightness decrement. Allows the selected remote input to be used to decrement the display
brightness. If no other remote input is set to
b.Inc
then once minimum brightness is reached the brightness
will start again from lowest highest brightness and decrement. If auto brightness is turned on then when
powered up the display will go to its auto brightness level and can be adjusted from there. If auto brightness
is turned off then the display will remember its last brightness setting on power up.
LD5RSMAN-2.0-0
63 of