![AMALGAMATED INSTRUMENT LD5-RS Operation And Instruction Manual Download Page 54](http://html1.mh-extra.com/html/amalgamated-instrument/ld5-rs/ld5-rs_operation-and-instruction-manual_2915959054.webp)
CH 0
- instructs the scanned instrument to return its channel zero display value.
CH 1
to
CH 8
) - instructs the scanned instrument to return its channel one or two etc. up to channel eight display
value.
AbUS
- if communicating with an instrument capable of AIC ABUS communication it is possible to use this option
to request a particular value by sending the ABUS value e.g. 800 hex is the ABUS value required to request the
primary display value from the instrument being scanned.
5.76
ABUS command value
Section:
CH 1
to
CH 8
Display:
AbuS CodE
Range:
0000
to
FFFF
Default Value:
Default Access Level
CAL
Function number
4dF0
to
4dF7
Seen when the
CONF CodE
function is set to
SCAN
. This function sets the ABUS command used when
Cmd
is
set to
AbUS
e.g.
0800
is the command for the primary display value. Thousands of commands are available some
of which are not useful in this software. A complete list is not available, contact supplier if a different command is
required.
5.77
Channel decimal point
Section:
CH 1
to
CH 8
Display:
d.Pnt
Range:
0
,
0.1
,
0.02
,
0.003
Default Value:
0
Default Access Level
CAL
Function number
4100
to
4107
Seen when the
CONF CodE
function is set to
SCAN
. This function sets the decimal point placement for the
selected channel.
5.78
Channel display rounding
Section:
CH 1
to
CH 8
Display:
d.rnd
Range:
1
to
5000
Default Value:
1
Default Access Level
CAL
Function number
43b0
to
43b7
Seen when the
CONF CodE
function is set to
SCAN
. This function sets the display rounding for the selected
channel. e.g. if set to
0.2
the display will only show multiples of 0.2.
54 of
LD5RSMAN-2.0-0