![AMALGAMATED INSTRUMENT LD5-RS Operation And Instruction Manual Download Page 6](http://html1.mh-extra.com/html/amalgamated-instrument/ld5-rs/ld5-rs_operation-and-instruction-manual_2915959006.webp)
The remote input mode uses the same access method as the Easy mode but also requires that one of
the available remote inputs is set to
ACCS
and that the selected remote input is activated i.e. shorted
to GND. The default access for this level is
NONE
so the access level will need to be changed if
access via this method is required.
P
F
Press and
hold F for
approx.
3 seconds
Also requires that the selected
remote input is set to
ACCSS
and is activated.
Remote input
mode
•
PIN 1 mode - Allows access to the level set by the
USr.1 LEUL
function in the
ACCS
menu.
The PIN 1 mode requires the
F
button is pressed and released then within 2 seconds press the
^
and
v
buttons at the same time. The PIN can be set via the
Pin.1 CodE
function in the
ACCS
menu. A
USr.1 LEUL
setting of
0
disables the PIN which means that there is no need to enter
the PIN. If the
USr.1 LEUL
function has been set to a number other than
NONE
then the first
function seen when entering via PIN 1 mode will be the function
CodE
. When this function is
seen the PIN value set at the
USr.1 LEUL
function must be entered via the
^
or
v
pushbuttons
followed by pressing
F
to accept the PIN before the user can progress to the setup functions.
P
P
F
F
Press and
release F then:
If a PIN has been set the
message
Code
will be seen.
Use
^
or
v
to enter the PIN
then press
F
to accept the PIN.
Press and
release Up
and Down
simultaneously
PIN 1
and
PIN 2
mode
•
PIN 2 mode - Allows access to the level set by the
USr.2 LEUL
function in the
ACCS
menu.
This method uses the same access method as PIN 1 mode above. A
Pin.2CodE
setting of
NonE
disables the PIN. If the
USr.1 LEUL
or a
USr.2 LEUL
function has been set to a number other
than
NonE
then the first function seen when entering via PIN 1/PIN2 mode will be the function
CodE
. When this function is seen the PIN value set at the
USr.1 LEUL
function can be entered
for access to the level set at the
Pin.1 ACCS
function or enter the
USr.2 LEUL
PIN to gain
access to the level set at the
Pin.2 ACCS
function. A correct code will allow access to the
functions at the selected level. An incorrect code will result in the
FUNC End
message being
seen indicating that access to setup functions has been refused and the display will return to normal
measurement mode.
•
Super Cal mode - This method can be used to gain access to all functions. If a PIN has been set and
forgotten use this method to access the PIN functions to check the settings. To access via Super Cal
mode with the instrument switched off hold in the
F
button whilst the instrument powers up. Keep
the button pressed until the
S.CAL
message is seen, you can then release the
F
button. Next press
and release
F
then within 2 seconds press and release the
^
and
v
pushbuttons simultaneously.
6 of