5.20
Serial input timeout alarm
Section:
AL 1
to
AL 8
Display:
tout
Range:
OFF
or
ON
Default Value:
OFF
Default Access Level
4
Function number
41d0
to
41d7
Allows the selected relay to be used to give an alarm indication if the serial input string is not received. Note that
this can be used in addition to the high and low setpoints.
5.21
Relay PI control functions
See the ”LD5 Series 8 Channel Scanning Monitor Output Addendum” booklet for details of the PI control
functions available
5.22
Relay PI control span
Section:
AL 1
to
AL 8
Display:
SPAN
Range:
Any display value
Default Value:
1000
Default Access Level
4
Function number
4290
to
4297
Allows setting of the control span, refer to “Setting up the relay PI control” chapter in the Addendum booklet.
5.23
Relay PI control setpoint
Section:
AL 1
to
AL 8
Display:
SEtP
Range:
Any display value
Default Value:
1000
Default Access Level
4
Function number
4200
to
4207
Allows setting of the control setpoint, refer to “Setting up the relay PI control” chapter in the Addendum booklet.
5.24
Relay PI control proportional gain value
Section:
AL 1
to
AL 8
Display:
P.g
Range:
Any display value
Default Value:
0.010
Default Access Level
4
Function number
4210
to
4217
Allow the relay PI control proportional gain to be set, refer to “Setting up the relay PI control” chapter in the
Addendum booklet.
LD5RSMAN-2.0-0
37 of