Amada MG3 Hot Bar Operating Instructions Manual Download Page 38

 

 

MG3 Hot Bar 

How to …? 

76609576EN-MG3-HB-V1_0a.docx 38/49 
 

 

5.11

 

Print Measured Values 

 

Here you can adjust the settings for the printing mode.  

 

A printer can be connected to the serial interface X51 (RS232). 

 

You can choose different values for channel 1 and 2. 

 

 

 

 

 

Go to screen PRINTER via MENU – PRINTER. 

 

 

Choose any number of values to be printed. 

 

BAUDRATE RS232 

Select the baud rate for the RS232 interface. 
 

MODE 
Select the printing mode. 

 

 

Possible printing modes are: 

 

off  

 

function  deactivated 

 

all  

 

all measurements 

 

reject  

rejected 

measurements 

 

prewarning 

measurements with active prewarning 

 

Interval  

measurements 

with an interval from 1..9999 

 

Interval/reject measurements 

with an interval from 1..9999 

   

 

and rejected measurements 

 

 

FORMAT 
Set to 

numeric

 to transmit the data to the printer (

graphic

 is only for 

internal use). 

  

 

PRINT-TEST 
Test the connection. 

 

 

 

 

Summary of Contents for MG3 Hot Bar

Page 1: ...Original Instructions 76609576EN MG3 HB V1 0a Digital Measuring Unit MG3 Hot Bar OPERATING INSTRUCTIONS...

Page 2: ...es Copyright 2015 AMADA MIYACHI EUROPE GmbH Lindberghstr 1 82178 Puchheim Germany All rights reserved The products named in this manual are only used for identification purposes and may be trademarks...

Page 3: ...232 Baud rate 14 3 1 2 6 Connector X60 and X61 CANopen 15 3 1 3 Connecting Temperature Sensor 15 3 1 4 Connecting Displacement Sensor optional 15 3 1 5 Connecting Force or Pressure Sensor optional 15...

Page 4: ...d Change Password 39 5 13 Configure the Logger 40 5 14 Display the Parameter Overview 42 5 15 Analyze Accept Reject Ratio 43 5 16 Monitor Statistical Values 44 5 17 Determine Limit Values Statisticall...

Page 5: ...with the unit This applies in particular to the notes on safety notes concerning safety regulations as well as safety and warning notes The fundamental principle applies Safety first Always carry out...

Page 6: ...on health or considerable material damage Warning This symbol warns of hazards emanating from electrical voltage Failure to heed these instructions can lead to grave irreversible or even life threate...

Page 7: ...a docx 7 49 1 3 2 Further Symbols Wear eye protection Disconnect power plug Wear protection gloves Pay attention to Operating Instructions Information This symbol indicates operator tips and other use...

Page 8: ...tions 2 temperatures with temperature peak temperature integral temperature average temperature gradient temperature difference heat time s Optional displacement and force or pressure Counter Up and d...

Page 9: ...with an adjustable support stand To have a non dazzling reading of the screen adjust the tilt angle with this stand Fig 1 MG3 with support stand Remove the MG3 Hot Bar from the packing and set it up a...

Page 10: ...code accordingly 1 99 X55 Control inputs from peripheral equipment X54 Control outputs to peripheral equipment X38 Inputs and outputs extended I O X31 Connector for displacement measuring or WK Adapte...

Page 11: ...r reached final value until counter reset 7 REJECT 2 Signal waiting as soon as monitored parameter is out of limits 8 Temperature 2 Signalizes the superior control that measurement is started by an ex...

Page 12: ...High measured values of a succeeding welding are not accepted With falling edge measured values are read in again starting with a new welding 13 Trigger 1 High if no measuring signal is detected a no...

Page 13: ...cutoff value is exceeded Low cutoff value is exceeded 18 21 Plus terminal Common plus terminal for PIN 14 13 11 12 10 9 8 7 6 5 4 3 30 29 28 27 26 25 24 23 19 Common 1 1 Common plus terminal 10 30 V f...

Page 14: ...e with falling edge 46 Null Sinking 1 Triggers with rising edge cooling time with falling edge 47 Parts Detection 2 Set parts detection HWScan Measured value is evaluated and part min 2 max 2 is set 4...

Page 15: ...ng kit For the mounting of the displacement sensor on the Hot Bar head please see the operating instructions of the Hot Bar system Connect the plug of the displacement sensor with the supplied sensor...

Page 16: ...er with the corresponding socket X31 on the rear side of the MG3 see chap 3 1 1 Note Switch off the MG3 Hot Bar before connecting 3 2 Initial Commissioning 3 2 1 Preconditions Check whether the MG3 Ho...

Page 17: ...box quadrant etc Grayed items are disabled and cannot be selected Input fields are Buttons which trigger a certain action when clicked Check boxes which have two switching conditions black on active g...

Page 18: ...ts USB USB stick missing blinking or access to USB stick on steady 4 3 Screen Display Fig 5 Screen Display 1 Display window quadrant 1 Q1 2 Display window quadrant 2 Q2 3 Menu bar 4 Program number sel...

Page 19: ...ue is outside the set prewarn limits red the measured value is outside the set limits gray invalid measurement range overflow 4 3 2 Menu Bar The menu bar comprises the following controls Display setti...

Page 20: ...he notebook or PC Press the remove USB stick button on the front of the MG3 Hot Bar before you remove the USB stick see chap 4 fig 4 Caution 4 5 UNDO Function The menu bar offers the button UNDO for a...

Page 21: ...1 Make General Settings After starting up and initialization the MG3 Hot Bar is in measuring mode Multiple Measurement RUN First you have to make the basic settings Click button MENU to change to SETT...

Page 22: ...activate this feature in order to speed up the program changeover Also deactivate this feature if you configure quadrant windows for SPC see chap 5 4 These SPC quadrant windows have to be configured t...

Page 23: ...EM screen displays general system information E g about MAINBOARD RAM assignment built in plug in cards hardware version software version PPC and DSP DISPLAY RAM assignment USB stick memory capacity d...

Page 24: ...s started when the force pressure set in the field TRIGGER is reached 0 displacement triggers if a signal is active at input X38 Zero penetration 1 or 2 TRIGGER Set the temperature force or pressure v...

Page 25: ...49 TEMPERATURE 1 2 Set the used type of thermo couples type K or type J EVALUATION WINDOW Activate and set a time slot MIN MAX in which measuring value monitoring shall happen Setting range 0 to 1030...

Page 26: ...gged at every valid triggering To switch to single measurement mode click button SINGLE The last box in the menu bar changes to SINGLE and the MG3 Hot Bar is now in single measurement mode To switch b...

Page 27: ...elect how the measured variable shall be displayed in this quadrant numeric graphic here you can also select the color of the graphs SPC Chart SPC Values SPC Statistical Process Control Select the mea...

Page 28: ...3 HB V1_0a docx 28 49 Button Click here to switch to the next quadrant SHOW LIMITS Activate this check box to show the set limits VIEW IN STATUS LINE CHANNEL 1 2 Choose heat time or program name to be...

Page 29: ...the main function DISTANCE SENSOR Choose the signal form and the resolution of the displacement sensor connected In case of digital displacement sensor you can also choose the maximum measurable stro...

Page 30: ...range 0 9999 sec COOL Set a minimum waiting time cool down time before the displacement value S3 S6 is calculated Setting range 0 9999 sec 5 5 2 Configure Part Detection CHANNEL Shows the currently s...

Page 31: ...way the welding head actually covers reference the distance between both electrodes is determined Note reference should only be used with part detection REFERENCE MEASURING Enabling the button will st...

Page 32: ...aximum values for the pressure or force If both values are zero the monitoring is deactivated SENSOR STANDARD 2P MODE Change between Standard and 2P Mode explanation see CALIBRATION Pressure sensor A...

Page 33: ...e value is shown at the load cell Set this value in the field VALUE1 resp VALUE2 Now deactivate CALIBRATION 5 7 Set Limit Values Go to screen LIMITS via MENU LIMITS Select up to two indicators to be m...

Page 34: ...ms and parameters You can choose between 4 different modes parameters save all program parameters PROG 1 99 and all global parameters parameters env see above and additionally all reference and envelo...

Page 35: ...set prewarn value is reached a signal is set at the corresponding output When the end value counting direction or zero counting direction is reached a signal is set at the corresponding output Go to s...

Page 36: ...in order to activate the view zoom and save menu Now the arrow pointing down right beside the check box is activated or arrows Click on the arrow pointing down right beside the check box to move one l...

Page 37: ...get back to previous menus Cursor menu Click to change to CURSOR menu Choose one of the four measured variables by clicking on it A crosshairs cursor appears in the color of the measured variable You...

Page 38: ...s to be printed BAUDRATE RS232 Select the baud rate for the RS232 interface MODE Select the printing mode Possible printing modes are off function deactivated all all measurements reject rejected meas...

Page 39: ...CHANGE Change the password only possible after correct login STATE Indicates whether you are authorized for access or not Note Password on delivery 0000 If you don t remember your password please con...

Page 40: ...nge it QUERY Choose the display mode all measurements rejected measurements last measurement detailed view CH Choose between channel 1 2 or 1 2 1 2 only for last measurement mode VALUE Select the valu...

Page 41: ...surements per channel Then the buffer is saved in a log file on the USB stick and starts new The log file has the file suffix log and is csv compatible time date counter number program number measured...

Page 42: ...An overview of the most important parameters for the selected program channel is displayed CHANNEL 1 Change to channel 2 Click again to change back to channel 1 SAVE Save the parameters as text file o...

Page 43: ...NALYSIS in the menu bar RESET Reset all counters for the selected channel DETAILS Select for which measured value the accept reject ratio shall be displayed separately LIMIT Limit value monitoring for...

Page 44: ...luation ends when this quadrant is reconfigured The SPC chart displays and calculates within sequential measurement intervals Procedure Configure a quadrant for the SPC chart mode see chap 5 4 Click i...

Page 45: ...recise the result of the limit value determination will be The histogram shows average and standard deviation values calculated of all measured values Click HISTOGRAM on the screen SPC chart DEFINE LI...

Page 46: ...front of the MG3 see chap 4 fig 4 It is automatically saved on the USB stick as a bmp file with the prefix SSH_xxx xxx means a consecutive number The LED on the USB stick illuminates for a few second...

Page 47: ...level stored program controller SPC There must be no triggering as long as the ready signal is low measuring interval During this measuring interval data is calculated compared and transferred to the...

Page 48: ...DATA CACHE FAILED 4020 CAN TIME OUT READ DATA Data transfer overloaded possible causes measuring interval too short too intensive rotary transducer Actuation during highly frequented measurements 4021...

Page 49: ...Display the Parameter Overview 42 Display Windows Quadrant 1 to 4 19 E Error messages 48 F Further Symbols 7 G General Information 5 General settings 21 H How to 21 I Initial Commissioning 16 Integrat...

Reviews: