Advanced parameters and commands
Pearleye Technical Manual
V2.2.0
61
Example
Set the factory default (115200 Baud and activate serial interface via Gigabit
Ethernet).
>
s=2A[CR]
s 0
Operation Mode
Integrator / Image
Store
F
E
D
C
B
A
9
8
7
6
5
4
3
2
1
0
-
-
-
-
-
-
-
-
e
a
a
-
x
x
x
x
rw
rw rw
rw rw rw rw
Bit group
Value (hex)
Description
x
O
110 Baud
1
300 Baud
2
600 Baud
3
1200 Baud
4
2400 Baud
5
4800 Baud
6
9600 Baud
7
19200 Baud
8
38400 Baud
9
57600 Baud
A
115200 Baud
a
00
RS232 via 15 pin SUB-D only.
01
Additionally activate second channel via Pleora device port serial 0.
10
Additionally activate second channel vis Pleora device port bulk 0
(MODE:UART).
11
Reserved.
e
0
Each character received from the host is echoed back to it.
1
No Echo.
Table 30: Baud rates: values