Introduction
Pearleye Technical Manual
V2.2.0
12
Caution
Environmental conditions
Operate the camera in dry and dust free environment. The
working temperature range (temperature of the camera
housing) depends on the camera model:
•
Pearleye P-007 LWIR: 25°C ... 45°C
•
Pearleye P-030 LWIR: 15 °C ... 35 °C
Only in this temperature range, the camera reaches optimal
image quality. Outside the optimum temperature range, the
image may look oversaturated..
Regarding the signal quality of the camera it is an advantage to
operate the camera under constant ambient air temperature.
This can e.g. be facilitated by assembling the camera on a
bigger metal corpus or by an additional active temperature
regulation.
The red LED at the backside of the camera (L3) or the value of
the register
T
serves for examination. If L3 permanently
lights up or rather the lowest bit of
T
is set, the camera
housing has a temperature which is outside of the optimum. In
this case provide an additional heating or cooling.
Note
Warm-up period
The warm-up of the camera takes 20 to 30 minutes.
To achive best performance and accurate temperature
measurements, wait until the warm-up is completed.