Advanced parameters and commands
Pearleye Technical Manual
V2.2.0
47
Advanced parameters and commands
This chapter describes the advanced configuration of the
Pearleye P-007 LWIR
/
Pearleye P-030 LWIR
models.
Two-point correction (A, B, E, J, K, N)
To configure the two-point correction, the following parameters:
The parameter
E
controls the operation mode of the two-point correction. The
following value assignments are valid:
Note
In most cases the factory setup of the camera is sufficient.
Only in special cases you need the advanced parameters and
commands.
Parameter Description
A
File number of the first correction image recorded at
cold
reference temperature.
B
File number of the first correction image recorded at
warm
reference temperature.
E
Two-point correction operating mode
J
Set value of the correction image of the actually chosen
correction data set recorded at
cold
reference temperature.
K
Set value of the correction image of the actually chosen
correction data set recorded at
warm
reference temperature.
N
File number for the set values of all correction data sets.
Table 16: Advanced parameters: two-point correction
Value
Description
0
Deactivated
Data is passed through transparently.
1
Activated
Correction data
A
and
B
with the set values
J
and
K
of the
actual correction memory page
S
is used.
Table 17: Parameter
E
: values