9/27
Edition 03 - May 97
C 50
Alcatel Vacuum Technology France - ASM 180 TD/TD+ - ASM 181 TD+ User’s Manual
Alphanumeric Control and Display
Panel (ACDP) operation
Operations available
during manual mode
test cycle
Display the time elapsed
since start of cycle
Press or
Print the current
measurement
Press
and
simultaneously
Switch off the filament
(see
C20
)
Press
Switch on the filament
(see
C20
)
Press
Modify the FIXED manual
reject setpoint :
Press on
.
Adjust the displayed setpoint
with the key
and
.
Press on
to valid and to
return to display during the
manual cycle test.
F
▲
▼
F
▲
▼
▲
▼
Manual test basic display
Timer: time elapsed since start of
cycle.
Instantaneous print-out of ticket if
the external printer is connected to
the printer interface (see
B40
) and
activated (see
page 5
)
(sample print-out see
page 10
)
Vacuum test only.
The filament indicator light
goes off.
Vacuum test only
The filament indicator light is lit
after a few seconds.
Only if FIXED reject is selected
and when leak detector is in test
mode.
The FIXED setpoint is memorised
until next modification.
It is active whichever the test status
(GL-FL or LDS).
Return to manual
test basic display
MAN./FL : 1.0 E-09
INLET P. : 1.0 E-03
MAN./FL : 1.0 E-09
INLET P. : 1.0 E-03
MAN./FL : 1.0 E-09
REJECT ? : 1.0 E-08
MAN./FL : 1.0 E-09
CHRONO : 27S
FIL/DEF. : 1.0 E-12
SPECTRO RESET !
MAN./FL : 1.0 E-09
CHRONO : 37S
F